CL-10 Plus – User Manual
40 / 75
ECPUS vers 0.2 eng
-
Select
Add method
again and fill in the settings as in Figure 3
-
A third method is added. This gives the information for calculate results in U/g Hb.
Figure 3
-
After the third method has been added, select
Save
, so that the changes are saved on a
new file, leaving unchanged the original file (
G6PD_R1.MD)
; the method, with the new
settings, will be saved as reported in the
Short Name
field
(G6PD).
-
Select
Edit
method
and rename the new file
G6PD
as
G6PD_HbR1.MD
(right hand button of the mouse placed on the document: Rename command)
.
-
From
File
menu enter
Select method
and choose
G6PD_HbR1.MD.
Start the working
session, following instructions on the package insert.
-
At the start of each measurement (
GO
command or F5 function key), it will appear a
window (Figure 4) in the same window as for the
sample id
and
sample name
, called Hb
g% value, where it is supposed to be added the haemoglobin value for that sample. This
value must be manually introduced by the operator at this stage for each sample, in order