Biochrom Anthos Fluido-2 Microplate Washer v2.0
Page 64
<+> or <->
Use these keys to move the manifold. The manifold can be moved in
steps of ~0.1mm (range: 0 – 18.8mm).
<enter>
Confirm the bottom dispense height and go to the next screen.
Screen 73. Overflow Dispensing Height
It is recommended to use a dispense volume slightly higher than the well
volume to produce an overflow phase. The slight flow rate variations
among dispensing needles are compensated by the overflow phase.
The overflow dispensing height defines the height of the manifold for
overflow washing. The position must be higher or equal as the aspiration
height. It may be defined by "teach-in" (move the manifold to the correct
position with the <+> and <-> keys).
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes.
Select “Mov” to position the manifold to the current overflow dispense
height and switch to the “teach-in” mode.
<+> or <->
Use these keys to enter the desired overflow dispense height in steps of
~0.1mm (range: 0 – 18.8mm).
<enter>
Confirm the overflow dispensing height and go to the next screen
.
Screen 74.
Overflow Dispense Height “Teach-in
’
mode
In this "teach-in" mode you can move the manifold to the correct height
with the <+> and <-> keys.
Function keys:
<F1> or <F2>
Select "
" to automatically detect the bottom of the well.
Ovflow disp ht:
Esc Mov 4.0mm
Ovflow disp ht:
0.1 mm