Biochrom Anthos Fluido-2 Microplate Washer v2.0
Page 62
<enter>
Confirm the aspiration height and go to the next screen.
Screen 69. Dispense Position
For washing the dispensing needle is placed above the well. From there it
dispenses the washing solution into the well. The aspiration needle
aspirates an overflow and prevents liquid spilling on next wells.
The dispense position defines the position of the plate below the manifold
during washing. The dispensing needle should be above the well and the
aspiration needle within. It may be defined by "teach-in" (move the plate
to the correct position with the <+> and <-> keys).
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes.
Select “Mov” to position the plate to the current dispense position and
switch to the “teach-in” mode.
<+> or <->
Use these keys to enter the dispense position in steps of 0.1mm (range:
-5.0 – 5.0mm).
<enter>
Confirm the dispense position and go to the next screen.
Screen 70.
Dispense Position “Teach-in
’
mode
In this "teach-in" mode the plate can be moved to the correct position with
the <+> and <-> keys. With “Mov” and “
” check if the needles are in
correct position.
Function keys:
<F1> or <F2>
Select "
" to automatically detect the bottom of the well (changes to
“
”).
Select "
" to move the manifold up.
Disp pos:
Esc Mov 1.6 mm
Disp pos:
Mov 0.1 mm