5
9.GUARANTEE
Six month.
10. COMMON TROUBLE
&
RESOLUTION
No. Common trouble
Reason
Solution
1
Connect the power,
but the
indicator
and nixie tube don’t
light
No 220V power
Check the power
Blown fuse
Replace
Power
switch
doesn’t turn on
Turn on power
switch
2
Set
speed with
actual
speed
are different
Voltage is on the
high side or low
side
Contact with
electric
department
3
Door locker can not
lock during run.
Door lock
Check and
Replace
4
There is noise after
starting
on
the
machine
Tube
with
liquid
not
equaling sample
or
rotor
nut
loosen
Make the tube
with
equaling
sample,
tighten the nut
5
Machine
didn’t
work according to
adjusted value
Forget to press
Ok
key
to
confirm
Reset value you
needed
and
press Ok key to
confirm
6
Time/Trouble
window show
E3
Door
switch
damaged or not
closed well
Press cover or
replace
E4
Cover is open
during running.
Closed
the
cover ,and start
up again
E5
Locked
driven
damaged
rotor,
plate
Replace motor
or driven plate.
Summary of Contents for LC-4KC
Page 1: ...Table top Low Speed Centrifuge LC 4KC LC 4KII User Manual BIOBASE GROUP Version 2020 11...
Page 7: ...6 11 CIRCUIT DIAGRAM...
Page 23: ...Table type Low Speed Centrifuge BKC TL5 BKC TL5II User Manual BIOBASE GROUP Version 2020 09...
Page 39: ...Table type Low Speed Centrifuge BKC TL6 BKC TL6II User Manual BIOBASE GROUP Version 2020 09...
Page 40: ......
Page 57: ...Table type High Speed Centrifuge BKC MH16 User Manual BIOBASE GROUP Version 2020 10...