Biobase LC-4KC User Manual Download Page 6

5

9.GUARANTEE

Six month.

10. COMMON TROUBLE

RESOLUTION

No. Common trouble

Reason

Solution

1

Connect the power,

but the

indicator

and nixie tube don’t

light

No 220V power

Check the power

Blown fuse

Replace

Power

switch

doesn’t turn on

Turn on power

switch

2

Set

speed with

actual

speed

are different

Voltage is on the

high side or low

side

Contact with

electric

department

3

Door locker can not

lock during run.

Door lock

Check and

Replace

4

There is noise after

starting

on

the

machine

Tube

with

liquid

not

equaling sample

or

rotor

nut

loosen

Make the tube

with

equaling

sample,

tighten the nut

5

Machine

didn’t

work according to

adjusted value

Forget to press

Ok

key

to

confirm

Reset value you

needed

and

press Ok key to

confirm

6

Time/Trouble

window show

E3

Door

switch

damaged or not

closed well

Press cover or

replace

E4

Cover is open

during running.

Closed

the

cover ,and start

up again

E5

Locked

driven

damaged

rotor,

plate

Replace motor

or driven plate.

Summary of Contents for LC-4KC

Page 1: ...Table top Low Speed Centrifuge LC 4KC LC 4KII User Manual BIOBASE GROUP Version 2020 11...

Page 2: ...RAL 2 2 PECULIARITY 2 3 PRINCIPLE OF OPERATION 2 4 SPECIFICATION 3 5 PARTS IDENTIFICATION 3 6 USING CONDITIONS 3 7 OPERATION PROCEDURE 3 8 MAINTAIN 4 9 GUARANTEE 5 10 COMMON TROUBLE RESOLUTION 5 11 CI...

Page 3: ...into a angle rotor symmetrically the rotor will spin on axis after you start the machine and the relative centrifuge force RCF comes into being while rotating It will make the composition at differen...

Page 4: ...system 4 Cover 5 Rotor 6 Nut 7 Motor 8 Toroidal transformer 9 Electric controlling system 10 Centrifugal chamber 11 Rubber foot PS there are two buttons on the right side The green button is power but...

Page 5: ...up to your setting time machine stop running If the machine is working do not open the cover Turn off power OFF press the safety lock and take out test tubes 8 MAINTAIN 8 1 Please read the manual boo...

Page 6: ...locker can not lock during run Door lock Check and Replace 4 There is noise after starting on the machine Tube with liquid not equaling sample or rotor nut loosen Make the tube with equaling sample ti...

Page 7: ...6 11 CIRCUIT DIAGRAM...

Page 8: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...

Page 9: ...Table type Low Speed Centrifuge BKC TL4 BKC TL4II BKC TL4III User Manual BIOBASE GROUP Version 2020 09...

Page 10: ...on principles 3 4 Technical parameters 4 5 The rotor parameters 4 6 The operation procedures 4 7 Safety Notice 10 8 Warranty Service Instruction 11 9 Comment Malfunctions and Obviating Methods 11 10 T...

Page 11: ...by the application of its special damper This machine also has a self balance function Step 2 Set up the rotor mode the rotating speed and the needed time The special PWM controlled speed control sys...

Page 12: ...rotating radius r the horizontal distance from the position of the test solutions to the axle center and the rotating speed n and here is the calculation formula RCF 1 118 10 6 n2 r g n The rotating s...

Page 13: ...e supply cord into the outlet and press down the power switch the power switch is on the left bottom of the machine Step 3 After pressing down the button of STOP OPEN put the rotors on the motor shaft...

Page 14: ...numbers 3 The operation icon 4 The setting of rotor modes 5 The setting of rotating speed 6 The SET button 7 The RCF button 8 The START button 9 The UP button 10 The DOWN button 11 The temperature set...

Page 15: ...er the first adjustment panel or touch the screen and click the rotating speed area to enter the same panel As it is shown below Press UP or DOWN button to adjust the rotating speed value Attention Th...

Page 16: ...ng the UP or DOWN button As it is shown below Press SET button to enter next step and the needed hours can be chosen by pressing the UP or DOWN button As it is shown below 3 On the bottom of the first...

Page 17: ...hird adjustment panel the adjustment panel of acceleration and deceleration or touch the screen and click the acceleration and deceleration area to enter the same panel As it is shown below Press the...

Page 18: ...eration pressing the STOP OPEN button can also stops the operation of the machine When the rotating speed reaches 0 the buzzer would make a buzzing sound and the cue words would be displayed on the wi...

Page 19: ...covered 5 The faceplate should not be uncovered in the situation that the motor and the rotors do not stop the operation completely 6 The power supply has its own earth wire so the null line and the e...

Page 20: ...f the rotors the machine would automatically run with the maximum rated speed of the rotors 6 If the operating temperature is too high and the cavity temperature is over 55 centigrade the machine cann...

Page 21: ...ing to the dampproof and quakeproof requirement the motor shaft should be painted a thin layer of lubricating grease and preserved in the well ventilated indoor places with the temperature ranging fro...

Page 22: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...

Page 23: ...Table type Low Speed Centrifuge BKC TL5 BKC TL5II User Manual BIOBASE GROUP Version 2020 09...

Page 24: ...on principles 3 4 Technical parameters 4 5 The rotor parameters 4 6 The operation procedures 5 7 Safety Notice 12 8 Warranty Service Instruction 12 9 Comment Malfunctions and Obviating Methods 13 10 T...

Page 25: ...y the application of its special damper This machine also has a self balance function Step 2 Set up the rotor mode the rotating speed and the needed time The special PWM controlled speed control syste...

Page 26: ...rotating radius r the horizontal distance from the position of the test solutions to the axle center and the rotating speed n and here is the calculation formula RCF 1 118 10 6n2r g n The rotating sp...

Page 27: ...100ml Time range 0 99h59min Rotate speed precision 50r min Noise 65dB Power AC 220V 50HZ 10A Dimension L W H 600 540 360mm Weight 35kg 5 The rotor parameters Model Product name Maximum capacity Maxim...

Page 28: ...avoided so as to protectthe flexible support system of the motor Then place the adapter with testtubes on the rotors and rotate the rotors The rotors should be turnedflexibly the adapter should be rot...

Page 29: ...rature and non refrigerated centrifuge has no such setting function 12 The display area of rotating speed and centrifugal force 13 The shift button of rotating speed and centrifugal force 14 The area...

Page 30: ...el As it is shown below Press UP or DOWN button to adjust the rotating speed value Attention The slowest rotating speed is 100r and the fastest is the maximum rated speed of the rotor 50r rotating spe...

Page 31: ...button and press the SET button again to enter the time adjustment area The needed minutes can be chosen by pressing the UP or DOWN button As it is shown below Press the SET button to enter next step...

Page 32: ...it is shown below Press the UP or DOWN button to choose the rotor number Attention Please take the real objects as the reference and find the corresponding real rotor pictures and rotor number 4 Pres...

Page 33: ...to preserve the adjusted value and then withdraw from the adjusted state Step7After checking out the settings mentioned above press the START button to start the operation of the machine During the o...

Page 34: ...completely the centrifugal process would restart by pressing and holding the IMPULS button Step10After the rotors stop the rotation completely press the STOP OPEN button to uncover the faceplate if th...

Page 35: ...c shock 7 After the whole separation the centrifuge machine should be cleaned up in time the power switch should be turn off and the power supply should be cut off 8 It should be very careful to take...

Page 36: ...ture is over 55 centigrade the machine cannot be operated After its cooling the machine goes back to work normally Besides the malfunctions mentioned above please contact the technicians of the manufa...

Page 37: ...ing to the dampproof and quakeproof requirement the motor shaft should be painted a thin layer of lubricating grease and preserved in the well ventilated indoor places with the temperature ranging fro...

Page 38: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...

Page 39: ...Table type Low Speed Centrifuge BKC TL6 BKC TL6II User Manual BIOBASE GROUP Version 2020 09...

Page 40: ......

Page 41: ...otor body or the hanging cup is cracked it is strictly forbidden to use otherwise it will burst Statement Jinan Biobase Biotech Co Ltd hereinafter referred to as Biobase has the final interpretation o...

Page 42: ...peration principles 3 4 Technical parameters 4 5 The rotor parameter 5 6 The operation procedures 6 7 Safety Notice 13 8 Warranty Service Instruction 13 9 Comment Malfunctions and Obviating Methods 14...

Page 43: ...The special PWM controlled speed control system assures its high control accuracy and its fast acceleration and deceleration function which provides different acceleration and deceleration time to be...

Page 44: ...ated by this formula 27 4 logeRmax logeRmin Ts min n2 r2 Type Rmax The rotating radius of the test solutions having the farthest distance with the axle center Rmin The rotating radius of the test solu...

Page 45: ...tor 12 10ml 6000 5140 16 15ml 5000 4400 24 10ml 4000 2810 4 250ml 4000 3290 4 50ml 5000 4600 4 100ml 5000 4800 Swing micro plate rotor 2 2 96 holes 4000 2360 BKC TL6 II Angle rotor 12 10ml 6000 5140 S...

Page 46: ...ace the adapter with testtubes on the rotors and rotate the rotors The rotors should be turnedflexibly the adapter should be rotated freely and there should be no block orcollision Step 4 Place the ce...

Page 47: ...The display area of rotating speed and centrifugal force 13 The shift button of rotating speed and centrifugal force 14 The area of acceleration and deceleration 15 The INPULS button 16 The Enter butt...

Page 48: ...100r and the fastest is the maximum rated speed of the rotor 50r rotating speed can be adjusted by each press Pressing the UP or DOWN button fixedly can adjust the speed value continuously As it is s...

Page 49: ...9 Press the SET button to enter next step and the needed hours can be chosen by pressing the UP or DOWN button As it is shown below...

Page 50: ...e real objects as the reference and find the corresponding real rotor pictures and rotor number 4 Press the SET button to enter the third adjustment panel the adjustment panel of acceleration and dece...

Page 51: ...below 5 When finishing every step mentioned above press the ENTER button for about 2 seconds to preserve the adjusted value and then withdraw from the adjusted state 7 After checking out the settings...

Page 52: ...IMPULS button the rotors begin the deceleration until they stop rotating During the deceleration process the IMPULS button cannot be pressed any longer when the rotors stop the rotation completely the...

Page 53: ...ic shock 7 After the whole separation the centrifuge machine should be cleaned up in time the power switch should be turn off and the power supply should be cut off 8 It should be very careful to take...

Page 54: ...ature is over 55 centigrade the machine cannot be operated After its cooling the machine goes back to work normally Besides the malfunctions mentioned above please contact the technicians of the manuf...

Page 55: ...g to the damp proof and quake proof requirement the motor shaft should be painted a thin layer of lubricating grease and preserved in the well ventilated indoor places with the temperature ranging fro...

Page 56: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...

Page 57: ...Table type High Speed Centrifuge BKC MH16 User Manual BIOBASE GROUP Version 2020 10...

Page 58: ...n principles 3 IV Technical parameters 4 V The rotor parameter 4 VI The operation procedures 5 VII Safety Notice 11 VIII Warranty Service Instruction 11 IX Comment Malfunctions and Obviating Methods 1...

Page 59: ...the application of its special damper This machine also has a self balance function Step 2 Set up the rotor mode the rotating speed and the needed time The special PWM controlled speed control system...

Page 60: ...e rotating radius r the horizontal distance from the position of the test solutions to the axle center and the rotating speed n and here is the calculation formula RCF 1 118 10 6n2r g n The rotating s...

Page 61: ...Table type high speed centrifuge Angle rotor 12 1 5 2 0ml 16500 Angle rotor 24 1 5 2 0ml 14000 Maximum rotate speed 16500r min Maximum centrifugal force 18780 g Maximum capacity 12 10ml Time range 0...

Page 62: ...p 4 Place the centrifuge tubes into the test tubes The centrifuge tubes should be placed symmetrically and the test solutions in the tubes should be even so as to make sure the equilibrium operation o...

Page 63: ...he SET button to enter the first adjustment panel or touch the screen and click the rotating speed area to enter the same panel As it is shown below Press the UP or DOWN button to adjust the rotating...

Page 64: ...N button and press the SET button again to enter the time adjustment area The needed minutes can be chosen by pressing the UP or DOWN button As it is shown below Press the SET button to enter next ste...

Page 65: ...it is shown below Press the UP or DOWN button to choose the rotor number Attention Please take the real objects as the reference and find the corresponding real rotor pictures and rotor number 4 Pres...

Page 66: ...current centrifugal force can be displayed on the window by pressing the RCF button or touching the screen and clicking the RCF icon Repress the button and the rotating speed value can be displayed ag...

Page 67: ...ld not be removed because of its high rebound force against the seal ring on top of the centrifugal chamber please press the faceplate slightly with one hand and repress the STOP OPEN button with the...

Page 68: ...c shock 7 After the whole separation the centrifuge machine should be cleaned up in time the power switch should be turn off and the power supply should be cut off 8 It should be very careful to take...

Page 69: ...ture is over 55 centigrade the machine cannot be operated After its cooling the machine goes back to work normally Besides the malfunctions mentioned above please contact the technicians of the manufa...

Page 70: ...ng to the dampproof and quakeproof requirement the motor shaft should be painted a thin layer of lubricating grease and preserved in the well ventilated indoor places with the temperature ranging from...

Page 71: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...