L.4. Shutdown for extended periods / for winter
L.4.1. Procedure
1)
Turn off the complete filtration equipment.
2)
Drain the pH injection heat contactor.
→
Never store the probe in distilled water.
→
Never touch the bulb of the probe.
3)
Equip yourself with:
-
the storage bottle initially mounted on the probe,
-
the stopper supplied,
-
a container filled with 10% hydrochloric acid (for the removal of scale),
-
a container filled with 2.6% bleach (for the removal of organic matter),
-
a suitable KCl solution (for the preservation of the probe).
4)
Disconnect the probe.
5)
Remove the probe from the probe holder.
6)
Soak the probe for 1 hour in the container of hydrochloric acid.
7)
Rinse the probe with tap water, without wiping it afterwards.
8)
Soak the probe for 1 hour in the container of bleach.
9)
Rinse the probe with tap water, without wiping it afterwards.
10)
Place the probe on a clean surface.
11)
Fill the storage vial with KCl solution.
12)
Insert the probe head into the storage vial.
13)
Tidy away and store the probe in a warm, dry place, positioning it vertically, with the bulb facing
downwards.
14)
Remove the nut from the probe holder and replace it with the stopper supplied, using Teflon tape
(not
supplied)
.
L.4.2. Notes to be taken into account
•
When the water temperature descends below 15 °C, the cabinet switch automatically into
“Low Temp
Mode
” (wintering mode) and, consequently, the electrolysis stops because, at this temperature, its
chemical properties cause premature wear to the electrolytic cell.
•
The water in the electrolytic cell must not freeze under any circumstances.
•
Provided the electrolytic cell is not connected to a power supply, leaving it installed on the pipework poses
no risk of damage.