3.5 Transfer Using Preprogrammed Protocols
All steps are identical to those in Sections 3.3 and 3.3.a, except for Section 3.3, step 2, which is
described here.
On the Home screen press the Navigation button that corresponds with LIST (Figure 12) to choose
between the Bio-Rad preprogrammed protocols and user-defined protocol lists (Figure 16).
Pressing the Bio-Rad preprogrammed protocols button accesses the protocols described in Table 3.
•
STANDARD SD
provides typical semi-dry transfer conditions for use with conventional semi-dry
western blotting consumables (see Section 3.6)
•
1.5 mm GEL
uses a longer transfer time (10 min) for more efficient transfer when using 1.5 mm thick
gels
•
HIGH MW
is optimized for more efficient transfer of large proteins (>150 kDa)
•
LOW MW
is optimized for more efficient transfer of small proteins (<30 kDa)
•
MIXED MW
is for efficient transfer of proteins over a broad range of molecular weights (5–150 kDa).
This protocol is also accessed via the Turbo Navigation button
•
Mini-TGX
is an ultrafast protocol that will transfer a single Mini-PROTEAN TGX gel with mixed MW
proteins (5–150 kDa) in 3 min with excellent efficiency.
Table 3. Bio-Rad preprogrammed protocols.
Fig. 16. Protocol options.
Protocol Name
MW (kDa)
Time (min)
2 Mini Format Gels or
1 Midi Format Gel
(per cassette)
1 Mini Format Gel
(per cassette)
Standard SD
Any
30
Up to 1.0 A; 25 V
1.5 mm GEL
Any
10
2.5 A; up to 25 V
1.3 A; up to 25 V
High MW
>150
10
Low MW
<30
5
Mixed MW (Turbo)
5–150
7
1 Mini-TGX
5–150
3
–
2.5 A; up to 25.0 V
Trans-Blot® Turbo™ Blotting System 13
Summary of Contents for Trans-Blot Turbo
Page 1: ...Trans Blot Turbo Blotting System Instruction Manual Catalog number 170 4155...
Page 2: ......
Page 4: ...ii Trans Blot Turbo Blotting System...
Page 6: ...iv Trans Blot Turbo Blotting System...
Page 8: ...vi Trans Blot Turbo Blotting System...
Page 12: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......