14 | QX200 Droplet Reader and QuantaSoft Software Instruction Manual
QX200 Droplet Reader and QuantaSoft Software Instruction Manual | 15
Chapter 3 Using QuantaSoft Software
Chapter 3 Using QuantaSoft Software
Fig. 15. Results table options.
Data from a CNV analysis are shown.
View table in graphical display window
(toggle)
Export data to .csv file
Select data from detector channel(s)
Toggle order in which channel data are displayed in the table
Display replicates separately, as merged wells (if Sample Name, Experiment, and
Assay Name and Type all match across the wells), or both
Status options:
OK
— automatic analysis successful
CHECK
—
automatic analysis unsuccessful; to view
concentration, use manual analysis tools
To get more details of an error, hover over
the check tab with the mouse arrow
Multi
— data automatically analyzed as part of a
multi-well selection
Manual
— droplets analyzed manually
Fig. 14. Data analysis interface.
Data from a CNV analysis are shown.
Graphical data
display
Results table
Well selector
Y-axis display options
Copy graph to clipboard or print
Note:
If a concentration estimate fails to appear in the concentration chart in QuantaSoft, this indicates the software
could not auto-analyze or assign droplets to positive or negative populations using its auto-analysis algorithm, or the well
had an unusually low droplet count (<10,000). Manually set a threshold and QuantaSoft Software will calculate a
concentration, which will appear in the concentration chart.