Table 8. (continued)
Temperature Mode Operation
Step
Procedure
Description
Table 9.
Examples of Power Requirements
The following table shows approximate power requirements to maintain 50°C using a Bio-Rad Sequi-Gen
®
sequencing apparatus with 0.4 mm spacer, 5% gel and 1xTBE buffer.
Gel
Dimensions
21 x 40 cm
21 x 50
21 x 80
38 x 50
Watts
80
100
140
120
11.
Select one of the following:
a.
TIME
b.
V-HOUR
c.
Press
to start the
run.
a. TIME: To specify automatic time control. You may then enter
a time from 1 minute to 99 hours and 59 minutes. After you
have entered a value, press
to start the run.
b. V-HOUR: To specify automatic volt-hour control for the run.
You may then enter a value from 1 to 99,999 volt-hours. Volt-
hours is the integration of voltage as a function of time. By
programming the unit to run for a specified number of volt-
hours, you get more accurate run-to-run reproducibility. After
you have entered a value, press
to start the run.
c. Pressing
will start the run. The run will continue until you
stop it.
Note: The run may be stopped at any time using either of the fol-
lowing keys:
Stops the run and displays the end-of-run parameters.
Interrupts operation of the power supply until the
key
is pressed. During the pause, the LCD display shows the
status of the run at the time the key was pressed. If you
do not wish to continue the run, press the
key.
25
POWER
PAC
Temperature Mode Operation