Table 2. (continued)

Manual Mode Operation

Step

Procedure

Description

9.

Select one of the following:

a.

TIME

a.

TIME:  To specify automatic time control.  You may then
enter a time from 1 minute to 99 hours and 59 minutes.
After you have entered a value, press

to start the run.

b.

V-HOUR

b.

V-HOUR:  To specify automatic volt-hour control for the run.
You may then enter a value from 1 to 99,999 volt-hours.
Volt-hours is the integration of voltage as a function of time.
By programming the unit to run for a specified number of
volt-hours, you get more accurate run-to-run reproducibility.
After you have entered a value, press

to start the run.

c.

Press

to start 

c.

Pressing

will start the run.  The run will continue until 

the run.

you stop it, or the run is completed. 

Note:  The run may be stopped at any time using either of the
following keys:

Stops the run and displays the end-of-run parameters.

Interrupts operation of the power supply until the
key is pressed.  During the pause, the LCD display
shows the status of the run at the time the key was
pressed.  Run parameters cannot be changed during a
pause.  If you do not wish to continue the run, press the

key.

10

POWER

    

PAC

Manual Mode Operation

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for PowerPac 1000

Page 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Page 2: ...1000 165 5054 165 5055 PowerPac 3000 165 5056 165 5057 165 5058 165 5059 165 5060 POWER PAC 3000 POWER PAC 1000 0 EXIT CLR 8 5 2 7 4 1 9 6 3 POWER PAC 3000 Artisan Technology Group Quality Instrumenta...

Page 3: ...x B Warranty and Ordering Information 33 List of Figures 1 PowerPac 3000 Power Supply 4 2 PowerPac Rear Panel 7 3 Manual Mode Flow Chart 11 4 Create Method Flow Chart 15 5 Display while Programmed Met...

Page 4: ...ied for safety compliance for use with PowerPac power supplies Use of other plugs or banana jacks is done at the user s own risk and is not recommended by Bio Rad When inserting and removing the molde...

Page 5: ...manual This safety cer tification does not extend to electrophoresis cells or accessories which are not I E C 1010 certified even when connected to this power supply This instrument should not be modi...

Page 6: ...rPac 3000 Voltage output Adjustable from 25 to 3000 volts dc VDC in increments of 1 volt Current output Adjustable from 1 to 400 milliAmps mA in increments of 1 mA Power output Adjustable from 1 to 40...

Page 7: ...0 international safety certification Automatic detection of no load rapid change in resistance arcing short circuit and ground leakage Automatic completion if desired of a run interrupted by a power f...

Page 8: ...rrent and power at the time the key was pressed Power output is stopped The run parameters can not be changed during a pause To resume the run press the key Exit key Returns you to the main menu Clear...

Page 9: ...EAR PANEL Figure 2 shows the following 1 Fan vents for cooling the unit 2 AC power input connector 3 Temperature probe jack PowerPac 3000 only 4 RS 232 serial port PowerPac 3000 only This is reserved...

Page 10: ...ide labeled l on the switch Note that the copyright screen and the firmware ver sion number appear briefly when the unit is turned on b The main menu is displayed as shown below Press the appropriate...

Page 11: ...value using the keypad If you make a mistake when entering a value press the key and re enter the value PowerPac 1000 Constant voltage The default limits are 500 mA and 250 W Constant current The def...

Page 12: ...more accurate run to run reproducibility After you have entered a value press to start the run c Press to start c Pressing will start the run The run will continue until the run you stop it or the run...

Page 13: ...CONSTANT W TIME V HOUR Optional Or Or Or Manual Mode CONSTANT CONSTANT CONSTANT V mA W TIME V HOUR CONSTANT TIME V HOUR PRESS TO RUN PRESS TO RUN Optional V HOUR 00000 TIME 00 00 SET xxxx V xxx mA xxx...

Page 14: ...led l on the switch b The main menu is displayed as shown below Press the appropriate soft key In the event of a power failure all operating parameters including time are retained in memory When power...

Page 15: ...ower W as the constant value Use the keypad to enter a value To change these limits select the limit to be changed and enter a value using the numeric keypad PowerPac 1000 Constant voltage The default...

Page 16: ...g a number which is not listed overwrites an existing method To start the run using the method s new parameter values press Figure 5 shows the contents of the Run screen Figure 6 shows the contents of...

Page 17: ...ode Creating a Method Or POWER FAIL DETECT OFF CONSTANT CONSTANT CONSTANT V mA W CONSTANT xxx W TIME V HOUR xxxx V CONSTANT xxx mA xxxx V xxx mA xxx W TIME V HOUR CONSTANT TIME V HOUR SET xxxx V xxx m...

Page 18: ...Program menu lists the following CREATE EDIT RUN and VIEW After you select EDIT the list of available methods is displayed You must select either power voltage or current The pro grammed constant para...

Page 19: ...er values be sure to select the same method number as that which was select ed for editing Note that the available method numbers are listed To start the run using the method s new parameter values pr...

Page 20: ...re assumes you have set up the unit as discussed in Table 3 Preliminary Setup Procedure at the beginning of this chapter The Program menu lists the following CREATE EDIT RUN and VIEW After you select...

Page 21: ...he contents of the first step This procedure is discussed in Table 6 If you have not yet created a method or if the method you want to run requires modifications select either CREATE or EDIT Follow th...

Page 22: ...olt hours 3 The method M and step S that is currently running 4 Actual output voltage current power Note The constant parameter is highlighted Figure 5 Display while Programmed Method is in Progress 4...

Page 23: ...ayed as shown in figure 6 Legend 1 Final voltage current and power 2 Programmed voltage current and power 3 Final method M and step S number and elapsed time 4 Repeat the run using the same method num...

Page 24: ...emperature Refer to Table 8 for approximate power requirements for Sequi Gen sequencing cell There is no substitution for optimizing the power con ditions required for your particular application and...

Page 25: ...um grease on the probe before attaching If the probe becomes dislodged from the glass plate during a run the temperature control will not work properly To remove the probe from the glass plate simply...

Page 26: ...ture mode unless there is cross over Enter the watts required to maintain the gel at the programmed temperature If desired change the default voltage and current Voltage may be a value between 25 and...

Page 27: ...run You may then enter a value from 1 to 99 999 volt hours Volt hours is the integration of voltage as a function of time By programming the unit to run for a specified number of volt hours you get mo...

Page 28: ...T POWER mA W TIME V HOUR V SET TIME V HOUR TIME V HOUR Optional Or PRESS TO RUN Temperature Mode PRESS TO RUN POWER FAIL DETECT OFF 45 C 400 3000 CONSTANT 45 C xxx W PRESS TO RUN SET 45 C xxx W SET 45...

Page 29: ...ected during an electro RESISTANCE phoresis run This may be due to a leaking cell SYSTEM WAS SHUT DOWN Check the cell for leaks PRESS CE TO RESET Reset the power supply by pressing the CE key to retur...

Page 30: ...d the run was completed after power was RUN WAS COMPLETED restored PRESS CE TO RESET Reset the power supply by pressing the CE key to return to the last screen The power supply then is ready to run EE...

Page 31: ...there may be a firmware or hardware problem Call Bio Rad for assistance TEMPERATURE PROBE REMOVED Temperature probe was removed from the power supply during the run SYSTEM WAS SHUT DOWN Reconnect the...

Page 32: ...nals 4 pair recessed female banana jacks floating and wired in parallel Timer Control 1 minute to 99 hours and 58 minutes or 99 999 volt hours No Load Detection PowerPac 1000 58 A or greater than 6 me...

Page 33: ...00 V DC rms isolation from high voltage outputs to earth ground PowerPac 3000 Internal fusing on both hot and neutral 4000 V DC rms isolation from high voltage outputs to earth ground Auto power up af...

Page 34: ...ative humidity non condensing Dimensions 29 L x 28 W x 11 H cm unit is stackable Weight PowerPac 1000 3 1 kg PowerPac 3000 3 7 kg 32 POWER PAC Specifications Artisan Technology Group Quality Instrumen...

Page 35: ...enendaal Phone 0318 540666 Fax 0318 542216 New Zealand Bio Rad Laboratories Pty Ltd Unit 15 Poland Court 21 Poland Road P O Box 100 051 North Shore Mail Centre Glenfield Auckland 10 Phone 09 443 3099...

Page 36: ...5060 PowerPac 3000 with temperature probe 220 240 V PowerPac 1000 165 5054 PowerPac 1000 100 120 V 165 5055 PowerPac 1000 200 240 V PowerPac 300 165 5050 PowerPac 300 100 120 V 165 5051 PowerPac 300...

Page 37: ...Fx 91 124 6398115 Israel Ph 03 951 4124 Fx 03 951 4129 Italy Ph 34 91 590 5200 Fx 34 91 590 5211 Japan Ph 03 5811 6270 Fx 03 5811 6272 Korea Ph 82 2 3473 4460 Fx 82 2 3472 7003 Latin America Ph 305 89...

Page 38: ...36 POWER PAC Table of Contents Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 39: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Reviews: