Water leak from dishes or from central basin
1) Control that the outlet pipe is correctly positioned (the outlet connecter of the stainer must be situated over the outlet
connecter of the laboratory).
2) Control that the pressure reducer valve is correctly regulated.
3) Verify the seal of inlet/outlet connecters of the running water dishes.
4) Control that internal and external outlet pipes aren’t blocked. If necessary, effect the cleaning using bleach or similar
products.
5) Check the correct working of the emergency float.
16) REPLACING THE FUSES
If the instrument doesn’t start, check that the electric cable is properly connected, there is current and the two fuses
under the instrument’s socket aren’t burnt. If necessary, change them with fuses of identic value.
-
Fuses holder socket (N. 2 delayed fuses of 4A - T4AH250V).
Instructions for the replacement:
Switch off the appliance and unplug the cable from the socket.
Make a light pressure on the little plastic carter covering the fuses under the power supply input (if necessary using a
little screwdriver), change them and close the carter.