11) ORDINARY MAINTENANCE AND FILTERS REPLACEMENT
It is possible to page to menu maintenance touching key REGULATIONS AND MAINTENANCE in main menu.
Menu MAINTENANCE is divided into two distinct parts (see pictures below):
1) The first, MAINTENANCE USER, is accessible only through password and allows to qualified technicians to check all
the machine parameters.
2) The second, MAINTENANCE FILTERS, allows the operator to check and update the filters hourcounter in case of
filter’s replacement. To page to filters maintenance, touch key MAINTENANCE FILTERS.
It is possible to zero/modify the hourcounter by pressing on the shown value.
The instrument is provided with an activated charcoal cartridge filter specific for alcohols and xilol, placed in the back
side of the instrument. The panel is provided with a digital hourcounter preset at 600 hours (middle life of a filter). For
the visualization of the filters hourcounter see the preceding paragraph. When the hourcounter reaches the value of
zero hours, an acoustic and visual alarm enters into action in order to advise the operator that the filter must be
replaced.
In order to proceed to the replacement, unscrew the two knobs on the left side, remove the cover, extract the exhausted
filter and replace it with a new one.
Enter in menu MAINTENANCE FILTERS and update the filters hourcounter.
Important:
Before effecting the replacement, switch off the instrument and unplug the power supply cable from the socket.
In case of accidental infiltration of liquids or foreign bodies in the filter area, contact immediately the Technical
Assistance Service and don’t use the instrument.
Dispose the exhausted filter in compliance with the regulation in force.
Reporting alarm filter exausted
Remove the filter exhausted, replace it with a new and
set the new duration in timer
- screen regulations and maintenance