Setup instructions for the IC-2A and optional blender using hose kit #2005ICH
This hose kit is intended for use when cylinders are NOT used.
Mount the blender bracket
(pins down)
approximately 12" from the top of the pole.
Place the blender in the bracket.
Place the IC-2A in the top tilt bracket.
Use the elbows (3) in the following locations. Face the threads to the rear.
Logic supply input on the IC-2A.
Patient supply input on the IC-2A.
Blender output.
The enclosed high pressure wye is to be used to split the oxygen source before the blender
so one leg goes to the blender input and the other to the IC-2A logic input.
Connect the 10' oxygen supply hose to the input of the wye.
Connect the short hose leg of the wye to the blender oxygen input .
Connect the long hose leg of the wye to the elbow on the IC-2A logic supply input.
Connect the 10' air supply hose to the blender air input.
Connect one end of the 2' supply hose to the elbow on the blender output and the other
end to the elbow on the IC-2A patient supply input.