BLENDER SETUP INSTRUCTIONS WITH HEAVY DUTY STAND
Setup instructions for the IC-2A and optional blender using hose kit #2005ICHC
This hose kit is intended for use when cylinders are used with the IC-2A complete
package, Catalog #8001AMBC.
Prior to mounting the pole to the base, slip the cylinder brackets over the end of the pole.
Put the open bracket on first and then the one with the bars across the rings. The bars
should be down (closest to the end of the pole). Install the pole to the base. Secure the
pole to the base with the washer and nut.
Mount the blender bracket
(pins down)
approximately 12" from the top of the pole.
Place the blender in the bracket.
Place the IC-2A in the top tilt bracket.
Install the oxygen cylinder with regulator in the cylinder bracket to the right and with the
supply output fitting facing forward. If using Bio-Med's Deluxe Regulator, then the 50 PSI
Out fitting should face forward.
Install the air cylinder with regulator in the cylinder bracket to the left and with the supply
output fitting facing to the rear. If using Bio-Med's Deluxe Regulator, then the 50 psi out
fitting should face forward.
Use the elbows (3) in the following locations. Face the threads as indicated:
Logic supply input on the IC-2A- face threads to the rear
Patient supply input on the IC-2A- face threads to the rear
Blender output- face threads to outside of blender air fitting
The high pressure wye is to be used to split your oxygen source before the blender so the
short leg goes to the blender input and the long leg goes to the IC-2A logic input. Connect
one end of a 2' oxygen supply hose to the input of this wye and the other end to the output
fitting on the oxygen cylinder regulator.
Connect one end of the 2' air supply hose to the output fitting on the air cylinder regulator
and the other end to the blender air input.
Connect the remaining 2' supply hose from the elbow on the blender output to the IC-2A
patient supply input.
If using Bio-Med's Deluxe Regulators, use the 10' air & oxygen hoses and connect the 50
PSI Input fittings to corresponding wall sources, if desired.