C. SPECIFICATIONS
Gas Supply
Logic Oxygen:
Clean, dry, medical grade 50
±
5 psi (345 ±34.5
kPa ) 100% oxygen.
Patient Supply:
Clean, dry, oil free 50
±
5 psi (345 ±34.5 kPa )
patient breathing gas mixture.
Inspiratory Time:
Calibrated
2
from 0.5 to 4 seconds (variable to 45
seconds or more, uncalibrated, in the IMV
range).
Flow rate:
Calibrated from 0 to 75 LPM.
Maximum Pressure Setting:
Variable from 0 to 120
±
20 cmH
2
O. This is
dependent upon the type of exhalation valve
used. The IC-2A is calibrated using the Bio-Med
Devices disposable patient circuit and exhalation
valve, Part #8002A.
Pressure Gauge:
-10 to +120 cmH
2
O;
±
3% full-scale accuracy.
Failsafe Two-Way Relief Valve: (Installed internally) opens at pressures above
120
±
20 cmH
2
O or below -4
±
1 cmH
2
O.
PEEP/CPAP Range:
Variable from 0 to 25
±
5 cmH
2
O.
Logic Gas Consumption:
Approximately 12 LPM. Varies with control
settings. Higher pressure increases
consumption.
Weight:
4.1 kg (9 lbs.)
Physical Dimensions:
8.57 X 15.56 X 26.04 cm (3-1/8 x 6-1/8 x 10-1/4
in.)
Storage Temperature:
32° to 122°F (0°to 50°C)
Operating Temperature: 14° to 122°F (-10°to 50°C)
2
INSP. TIME and EXP. TIME control settings are calibrated at sea level and 20 degrees Celsius, using
USP oxygen. Large changes of barometric pressure or altitude changes, or use of diluted oxygen will
affect time calibration. Time settings are repeatable within 5%.