NCR-80 Four-wire 4-20 mA/HART
925-0332 Rev B
9 Diagnosis, asset management and service
57
The following table shows the error codes and text messages in the status message "Function
check" and provides information on causes as well as corrective measures.
The following table shows the error codes and text messages in the status message "Out of
specification" and provides information on causes as well as corrective measures.
Code
Text message
Cause
Rectification
DevSpec State in
CMD 48
F260
Error in the calibration
- Error in the calibration
carried out in the fac-
tory
- Error in the EEPROM
- Exchanging the elec-
tronics
- Send instrument for
Repair
Bit 8 of Byte 0…5
F261
Error in the instrument
settings
- Error during setup
- False signal suppres-
sion faulty
- Error when carrying out
a reset
- Repeat setup
- Carry out a reset
Bit 9 of Byte 0…5
F264
Installation/Setup error
- Adjustment not within
the vessel height/mea-
suring range
- Max. measuring range
of the instrument not
sufficient
- Check or correct instal-
lation and/or parameter
adjustment
- Use an instrument with
bigger measuring range
Bit 10 of Byte 0…5
F265
Measurement function
disturbed
- Sensor no longer car-
ries out a measurement
- Operating voltage too
low
- Check operating volt-
age
- Carry out a reset
- Disconnect operating
voltage briefly
Bit 11 of Byte 0…5
Code
Text message
Cause
Rectification
DevSpec State in
CMD 48
C700
Simulation active
- A simulation is active
- Finish simulation
- Wait for the automatic
end after 60 mins.
"
Simulation Active
" in
"
Standardized Status
0
"
Code
Text message
Cause
Rectification
DevSpec State in
CMD 48
S600
Impermissible elec-
tronics Temperature
- Temperature of the
electronics in the
non-specified range
- Check ambient tem-
perature
- Isolate electronics
- Use instrument with
higher temperature
range
Bit 8 of Byte 14…24
S601
Overfilling
- Danger of vessel over-
filling
- Make sure that there is
no further filling
- Check level in the ves-
sel
Bit 9 of Byte 14…24
S603
Impermissible operat-
ing Voltage
- Operating voltage be-
low specified range
- Check electrical con-
nection
- If necessary, increase
operating voltage
Bit 11 of Byte 14…24