CNCR-190 • Two-wire 4-20 mA/HART
15
Out of specification
Code
Text message
Cause
Rectification
DevSpec State
in CMD 48
S600
Impermissible
electronics temperature
Temperature of the
electronics out of specified
range
Check ambient temperature
Insulate electronics
Byte 23, Bit 4 of
Byte 14 to 24
S601
Overfilling
Danger of vessel overfilling
Make sure that there is no
further filling
Check level in the vessel
Byte 23, Bit 5 of
Byte 14 to 24
S603
Impermissible
operating voltage
Terminal voltage too low
Check terminal voltage, increase
operating voltage
Byte 23, Bit 6 of
Byte 14 to 24
Function check
Code
Text message
Cause
Rectification
DevSpec State
in CMD 48
C700
Simulation active
A simulation is active
Finish simulation
Wait for the automatic end
after 60 min.
“
Simulation
Active” in
“Standardized
Status 0”
Code
Text message
Cause
Rectification
F105
Determine
measured value
The instrument is still in the
boot up phase, the measured
value could not yet be
determined
Wait for the end of the boot up phase
Duration up to 3 minutes depending
on the measurement environment and
parameter settings
Byte 5, Byte 5,
Bit 6 of Byte 0
to 5
F260
Error in the calibration
Checksum error in the
calibration values
Error in the EEPROM
Send instrument in for repair
Byte 4, Bit 0
of Byte 0 to 5
F261
Error in the instrument
settings
Error during setup
False signal suppression
faulty
Error when carrying out a
reset
Repeat setup
Reset instrument
Byte 4, Bit 1
of Byte 0 to 5
F265
Measurement function
disturbed
Program sequence of the
measuring function disturbed
Device restarts automatically
Byte 4, Bit 3
of Byte 0 to 5
Failure