Program entry as set-point ramp (example)
On
Off
t/min.
Operation line 1 = air flap
W/°C
t/min.
01
02
03
04
0
50
100
150
200
250
30
120
180
380
05
Tolerance
minimum
Tolerance
maximum
Program table corresponding to the diagram above:
Program
section
Set-point
temp.
Fan
Section
time
Operation
line1
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01
50
100 %
00:30:00
Off
1
0
-1999
+9999
02
100
100 %
01:30:00
Off
1
0
-5
+5
03
200
100 %
01:00:00
Off
1
0
-2
+2
04
200
100 %
03:20:00
Off
1
0
-1999
+9999
05
100
100 %
00:00:01
Off
1
0
-1999
+9999
Now enter the values of the above program table into one of the 25 program places of the controller MB1:
PGM
08:43:55 15.02.02
PGM
08:43:55 15.12.13