M (E2) 04/2019
page 14/71
3.4
Location of installation and ambient conditions
Set up the chamber on a flat, even and non-flammable surface, free from vibration, and in a well-
ventilated, dry location and align it using a spirit level. The site of installation must be capable of support-
ing the chamber’s weight (see technical data, chap. 15.3). The chambers are designed for setting up
inside a building (indoor use).
CAUTION
Danger of overheating.
Damage to the chamber.
Do NOT set up chambers in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
•
Permissible ambient temperature range during operation: +18 °C up to +40 °C /
64.4 °F to 104 °F
. At
elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +25 °C /
77 °F
to which the specified technical data relate. In the case of different
ambient conditions, deviations from the indicated data are possible.
•
Permissible ambient humidity: 70 % r.H. max., non-condensing.
•
Installation height: max. 2000 m /
6562 ft.
above sea level.
When placing several chambers of the same size side by side, maintain a minimum distance of 250 mm /
9.84 in
between each chamber. Wall distances: rear 100 mm /
3.94 in
, sides 160 mm / 6
.29 in
.
CAUTION
Danger by stacking.
Damage to the chambers.
Do NOT place the chambers on top of each other.
To completely separate the chamber from the power supply, you must disconnect the power plug. Install
the unit in a way that the power plug is easily accessible and can be easily pulled in case of danger.
For the user there is no risk of temporary overvoltages in the sense of EN 61010-1:2010.
Do not install or operate the chambers in potentially explosive areas.
DANGER
Explosion hazard.
Danger of death.
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.
CAUTION
The exhaust duct will become hot during operation.
Danger of burning.
Do NOT touch the exhaust duct during operation.
Summary of Contents for M 115
Page 58: ...M E2 04 2019 page 58 71 15 7 Dimensions M 53...
Page 59: ...M E2 04 2019 page 59 71 15 8 Dimensions M 115...
Page 60: ...M E2 04 2019 page 60 71 15 9 Dimensions M 240...
Page 61: ...M E2 04 2019 page 61 71 15 10 Dimensions M 400...
Page 62: ...M E2 04 2019 page 62 71 15 11 Dimensions M 720...
Page 63: ...M E2 04 2019 page 63 71 16 EU Declaration of Conformity...
Page 64: ...M E2 04 2019 page 64 71...
Page 65: ...M E2 04 2019 page 65 71 17 Product registration...