M (E2) 04/2019
page 28/71
You can enter
Program sections
into this program table.
Hit the PGM button. An inquiry display appears allowing you to enter or delete individual program sec-
tions:
ZP-Abschnitt Abs. Nr. 5
insert
delete
In this view, new program lines can be entered or deleted:
new
New lines are added below in the table
insert
New lines are added above a previously selected line
delete
Individual lines that have been selected previously are deleted
Create as many lines, i.e. program sections, as desired. As a next step, values can be entered into these
lines. It is possible to add supplementary lines later or to delete individual lines at any time.
To enter values, select the corresponding line via the
arrow keys.
Hit the “ENTER” button. The
program editor
ap-
pears.
Enter the individual values of the selected program section.
Program editor Abs.Nr. 6
****.*
Setpoint 1
+100.0
FAN
Operating contacts
Time
00:45:00
Repeat Section
5
Repeat Number
10
Tol.-band min.
-1999.0
Tol.-band max.
+9999.0
Parameter set
1
--
Temperature value at the start of the program section
-- Fan speed in %
-- Operating contact (operation line) 1 = air flap open / closed
-- Duration of the program section
-- No. of start section in case of repeat cycles
-- No. of duplicates in case of repeat cycles
-- Temperature limits (maximum / minimum temperature).
In case
of exceeding: temporary program stop.
-- Pre-selected value (Do NOT change!)
Select the parameters via the arrow keys and confirm by pressing “ENTER”..
Then enter the values via the arrow keys, and confirm the entry by pressing “ENTER”..
Adapt the temperature safety device class 2 (chap. 10.1) or the temperature safety device
class 3.1 (option, chap. 10.2) to the highest temperature set-point value of the program actual-
ly used. Check the safety device for each temperature program and adapt it if necessary.
Set the set-point of temperature safety device class 2 or class 3.1 (option) by about 5 °C to 10
°C above the controller temperature set-point.
Performance after completing the program:
The controller changes to Idle Mode. The heating is inactive; the chamber approximates ambient temper-
ature. The fan turns at a 50% rate.
PGM
08:43:55 15.02.02
PGM
08:43:55 15.12.13
Summary of Contents for M 115
Page 58: ...M E2 04 2019 page 58 71 15 7 Dimensions M 53...
Page 59: ...M E2 04 2019 page 59 71 15 8 Dimensions M 115...
Page 60: ...M E2 04 2019 page 60 71 15 9 Dimensions M 240...
Page 61: ...M E2 04 2019 page 61 71 15 10 Dimensions M 400...
Page 62: ...M E2 04 2019 page 62 71 15 11 Dimensions M 720...
Page 63: ...M E2 04 2019 page 63 71 16 EU Declaration of Conformity...
Page 64: ...M E2 04 2019 page 64 71...
Page 65: ...M E2 04 2019 page 65 71 17 Product registration...