KBF-LQC (E5.2) 12/2010
page 87/106
We recommend taking out a maintenance agreement. Please consult BINDER Service:
BINDER telephone hotline:
+49 (0) 7462 2005 555
BINDER fax hotline:
+49 (0) 7462 2005 93555
BINDER e-mail hotline:
BINDER service hotline USA:
+1 866 885 9794 or +1 631 224 4340 (toll-free in the USA)
BINDER service hotline Spain
+34 9492 677 23
BINDER service hotline Asia Pacific:
+852 39070500 or +852 39070503
BINDER service hotline Russia and CIS
+7 495 98815 17
BINDER Internet website
http://www.binder-world.com
BINDER address
BINDER GmbH, post office box 102, D-78502 Tuttlingen
International customers, please contact your local BINDER distributor.
18.2 Replacement of the fluorescent tubes
The average life expectancy of the fluorescent tubes is about 10,000 hours. We recommend replacing
the tubes every year in order to ensure full light intensity.
To replace the fluorescent tubes, unscrew and remove the clamping strips resting against the glass plate
(Allen screwdriver). Then lift the plate from the cassette. Rotate the tubes by 90° and pull them out of
their holders. When setting in the new tubes, observe the tube orientation (inscription, Figure 26). Fix the
glass plate screwing the clamping strips.
Figure 39: Dismounting a light cassette
Always replace all the fluorescent tubes of a light cassette together. Otherwise, homogeneity
of light intensity cannot be guaranteed.
18.3 Calibrating the light sensors and adjusting the controller display
The light sensors are delivered with a calibration certificate on which at least 2 measuring values with the
corresponding measuring current values are indicated. So it is possible to adjust the input channels to the
sensor current values. Instead of the sensor a calibrating device is plugged in the sensor socket inside
the chamber. This source of current is voltage stabilized and independent from the power supply voltage.
The supply point for the sensors inside the chamber is automatically left out.