KBF-LQC (E5.2) 12/2010
page 13/106
The climatic chamber KBF-LQC does not dispose of any measures of explosion protection.
DANGER
Explosion hazard.
Danger of death.
∅
Do NOT introduce any substance into the climatic chamber which is combustible or
explosive at working temperature.
∅
NO explosive dust or air-solvent mixture in the inner chamber.
Any solvent contained in the charging material must not be explosive or inflammable. I.e., irrespective of
the solvent concentration in the steam room, NO explosive mixture with air must form. The temperature
inside the chamber must lie below the flash point or below the sublimation point of the charging material.
Familiarize yourself with the physical and chemical properties of the charging material, as well as the
contained moisture constituent and its behavior with the addition of heat energy and humidity.
Familiarize yourself with any potential health risks caused by the charging material, the contained mois-
ture constituent or by reaction products that may arise during the temperature process. Take adequate
measures to exclude such risks prior to putting the climatic chamber into operation.
DANGER
Electrical hazard.
Danger of death.
∅
The unit must NOT become wet during operation or maintenance.
The climatic chambers were produced in accordance with VDE regulations and were routinely tested in
accordance to VDE 0411-1 (IEC 61010-1).
CAUTION
The glass doors, the inner chamber, and the light cassettes will become hot during
operation.
Danger of burning.
∅
Do NOT touch the glass doors, the inner surfaces, the light cassettes or the charging
material during operation.
WARNING
Stability hazard.
Danger of injury.
Damage to the unit and the charging material.
Housing cover breakaway.
∅
Do NOT climb on the lower housing cover.
∅
Do NOT load the lower housing cover with heavy objects while the unit door is open.