KBF-LQC (E5.2) 12/2010
page 21/106
3.2
Guidelines for safe lifting and transportation
The front castors of the KBF-LQC can be blocked by brakes. Please move the units with castors only
when empty and on an even surface, otherwise the castors may be damaged. After operation please
observe the guidelines for temporarily decommissioning the unit (chap. 19.2).
CAUTION
Sliding or tilting of the unit.
Damage to the unit.
Risk of injury by lifting heavy loads.
¾
Transport the unit only in its original packaging.
¾
Secure the climatic chamber with transport straps for transport.
∅
Do NOT lift or transport the unit using the door handle, the door or the lower housing.
¾
Lift units size 240 corners with the aid of 6 people or with a fork lifter. Set the fork lifter
only from the front or rear in the middle of the unit.
¾
Lift units size 720 using technical devices (fork lifter). Set the fork lifter only from the
front or rear in the middle of the unit.
∅
Do NOT set the fork lifter from the unit side.
You can order transport packing for transportation purposes from BINDER service.
Permissible ambient temperature range during transport:
•
If the steam humidifying system has NOT been emptied: +3 °C /
37.4°F
to +60 °C /
140°F
.
•
After BINDER Service has emptied the steam humidifying system: -10 °C /
14°F
to +60 °C /
140°F
.
With temperatures below +3 °C /
37.4°F
, water must be completely removed from the humidifying sys-
tem.
CAUTION
Transport below +3 °C / 37.4°F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the unit.
¾
Contact BINDER Service before any transportation below +3 °C /
37.4°F
.
3.3 Storage
Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for tempo-
rary decommissioning (chap. 19.2).
Permissible ambient temperature range during storage:
•
If the steam humidifying system has NOT been emptied: +3 °C /
37.4°F
to +60 °C /
140°F
.
•
After BINDER Service has emptied the steam humidifying system: -10 °C /
14°F
to +60 °C /
140°F