1-72
D70E
If you have selected
Serial Number
from the
Setup Menu
the following data will be transmitted:
1st character
$(24H)
2nd character
M(4DH)
3rd character
P(50H)
4th - 11th character
Serial number
12th-18th character
MPP identification
code or indications on
status of scale:
NO<SP>STAB weight is not
stable NO<SP>VAL<SP>
weight is not valid
ERRMEM<SP> weight
memorisation error
19th - 26th character
Weight with sign and decimal
separator
27th - 28th character
<um>
Unit of measurement
29th - 30th character
<CRC>
String control field:
this is calculated with a XOR
(or exclusive) operation
involving all previous
characters. The value
calculated is shown as two
hexadecimal characters, e.g.
if the result is 62H, the two
checksum characters are "6"
(36H) and "2" (32H).
If you enabled the parameter "ABTA" in the string MPP selected it appears on
the tare value.
The 10 digits of tare (weight 8 + 2 UM) are queued to the weight.
Example.
If you have selected
Std
+ABTA=SI in the
Setup Menu
, the
following data will be transmitted:
1st character
$(24H)
Summary of Contents for D70E
Page 2: ......
Page 29: ...1 29 D70E 9 pin connector Maximum number of terminals that may be connected 16 ...
Page 80: ...1 80 D70E ...