Intended use
Product intended for professional use only. The iOptima
INT
systems are only intended
for use in general dentistry, restorative dentistry, oral surgery, implantology and
endodontics procedure by dentists and dental professionals in a dental office.
2.3
Notation
•
A
,
B
,
C
, etc.
Text preceded by a letter indicates a procedure to be carried out step-by-step.
•
(1)
,
(2)
,
(3)
, etc.
Text preceded by a number indicates text used in conjunction with an illustration.
3 Electromagnetic compatibility
for iOptima
INT
3.1
Electromagnetic compatibility warnings
The intended EM environment (per IEC 60601-1-2 ed. 4.0) is Professional
healthcare facility environment.
⚠
CAUTION
The iOptima
INT
complies with the EMC requirements according to IEC 60601-1-2.
Radio transmitting equipment, cellular phones, etc., should not be used in the
immediate vicinity of the device, since this could affect its operation. The device is
not suitable for being used close to high-frequency surgical equipment, magnetic
resonance imaging (MRI) and other similar devices where the intensity of
electromagnetic disturbances is high. In any case, ensure that no high frequency
cables are routed above or near the device. If in doubt, contact a qualified
technician or Bien-Air Dental SA.
Special precautions should be taken when using strong emission sources such as
high-frequency surgical equipment and other similar devices, to ensure that HF
cables are not routed above or near the device. If in doubt, please contact a
qualified technician or Bien-Air.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the iOptima
INT
, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
⚠
CAUTION
The use of accessories, transducers and cables other than those specified, with
the exception of transducers and cables sold by Bien-Air as spare parts for
internal components, may result in increased emissions or decreased immunity.