![Bien Air 1700704-001 Installation Instructions Manual Download Page 21](http://html1.mh-extra.com/html/bien-air/1700704-001/1700704-001_installation-instructions-manual_2749374021.webp)
ENG
17
I.
Connect the dual motor whitebox (1502569) to the motor hoses following the
color code on the connectors. MX2 Motor shall be connected to (1.1), MX-i motor
shall be connected to (1.2).
J.
Connect the Power supply reset switch to the DMX3 Whitebox (1601015) (2).
K.
Drill a hole of diameter 14 mm to 15 mm (9/16" to 19/32") in the unit and mount
MX-i Connector (1502529) with the dedicated tool (3600139) and the open-end
wrench:
• Insert the MX-i Connector (1502529) (1) from inside the dental unit.
• Add the nut (2) provided with the MX-i Connector (1502529) outside the
dental unit.
• Tighten the nut (3) using the tool (3600139) (4) and the open-end wrench.
FIG. 24
FIG. 25
FIG. 26
1.2
1.1
2
1
2
3
4