![Bien Air 1700704-001 Installation Instructions Manual Download Page 33](http://html1.mh-extra.com/html/bien-air/1700704-001/1700704-001_installation-instructions-manual_2749374033.webp)
ENG
29
7 Integration Checklist
Please go through the following checklist:
Note:
If there is any unchecked box remaining, please do not use the system and refer to Troubleshooting
procedures on the iOptima application.
•
For use on iPad mini®, check that the screen’s rotation lock is disabled and make sure the screen
displays the correct orientation according to its position (portrait or landscape orientation)
A.
CHECK that overall installation is conformed with the wiring diagram
B.
CHECK that all mechanical component are well fixed
•
Check that there is no mechanical traction on any wire
•
Check that each electrical element could not be harmed by dental unit surrounding
•
Check that USB Lightning cable is held by the nylon cable tie provided
•
Check that there is no mechanical traction on any pneumatic tube
•
Check that each pneumatic connection is correct with a light pull test
•
Check that there is no squeezing or deformation on any pneumatic tube
•
Check that all the connectors are correctly plugged in
•
Check that whiteboxes are closed and fixed with Velcro
•
Check that electronic boards are not able to move inside the whiteboxes
C.
CHECK that power supply is well installed
•
Only allow use of the power supply inside of the dental unit
•
Check that there is a circuit breaker upstream power supply
D.
CHECK that air flow system is well configured
•
Verify that an air filter is installed in the dental unit
•
Verify that an input air pressure is at least 3 bar / 43.5 psi and maximum 5bar / 72.5 psi
•
Check that the air flow rate at the motor nose is 10Nl/min (Normal Liter per minute)
•
Check that air is flowing from the motor nose in Resto mode and check that no air is flowing
from the motor nose in Endo mode
E.
CHECK that system is functional
•
Check that when pulling up the motor and pressing footpedal, that it turns ON
•
Check that motor LIGHT is ON when the motor is ON
•
Check that the rotation direction motor is clockwise when in RESTO
•
Check that the App configuration is consistent with the wiring (for example Motor1 = MX2
and Motor2 = MX2 in case of set 1700705)
•
Check on the iPod/iPad that the maximum speed reachable is 40’000rpm
•
Check that the power supply reset switch works
•
If installed, check that the peristaltic pump works; ON/OFF check, irrigation level, activation
(only in SR mode)
•
If installed, check that the scaler works properly; power output follows power level from the application
•
If installed, check that the scaler control mode is properly set