ESH 04
4
1
WARNINGS AND SAFETY INSTRUCTIONS
The emergency warmer can only be used or connected in vehicles
where the on-board installations comply with the standards and
regulations in force.
The emergency warmer must always be used in a hanging position.
The device shall not be used for infusion in a horizontal or an incorrect
position. This could lead to a hazardous situation due air bubbles
leaking into the infusion set.
Make sure there are no air bubbles in the tubing set in case the device
falls down or temporarily gets into an inclined position.
Always ensure that the infusion bottles are vertical aligned with
infusion outlet directed downwards and positioned 20cm higher than
the patient infusion access.
To minimize the risk of air bubbles entering the tubing set, always use
an infusion set with drip chamber or air trap
The treating physician must check the condition of the patient at
regular intervals. Failure to do so could seriously harm the patient.
If the device is attached to infusion stands or tripods, then these must
be sufficiently stable and able to carry the load.
The device must only be used with electrical installations that comply
with the standards and regulations in force.
Only use original BIEGLER parts in conjunction with the ESH 04:
mains adaptor 30 W E2EFMW3, PowerPort 12 V.
Unplugging the mains plug is the only safe way to disconnect from the
mains power supply. Only plug the mains adaptor into easily
accessible sockets, to ensure that it can be unplugged from the mains
quickly if required.
The device must not be used if the power connector is damaged.
The device must not be used in rooms with potential explosion hazard.
Persons and services authorized by BIEGLER must carry out repairs
and modifications on the device.
The device must not be immersed in liquids or sterilized with steam or
by thermochemical methods.