ESH 04
10
Note: The safety button must always be pressed simultaneously along
with the desired function to prevent only one of the buttons being
pressed inadvertently.
Starting up for use
Press the buttons and simultaneously to switch on the device
(green ON LED lights up). When switched on and running on battery
power, the infusion solutions are heated to 37 °C or kept warm for
approx. 90 minutes at 37 ± 3 °C.
The yellow LED on the left indicates the status of the battery pack.
A
flashing LED
indicates that the
battery is empty.
Setting up for use
Release pressure from the pressure cuff completely.
Hang the emergency warmer at an appropriate place.
Connect the infusion fittings.
If ordered by the physician, the pressure cuff can be used for pressure
infusion.
Attention: The emergency warmer must not be used in a lying or
upside-down position. If used improperly, especially in case of
pressure infusion, there is the risk of air bubbles entering the
infusion line. This can lead to serious damage to the patient
(danger of air embolism).
Decommissioning
Switch off the device via the Standby button and if no charging is
required, unplug the mains adaptor.