24
switch Test - Adjust End point
To Adjust the End point temperature press
the Set button four times and continue to
hold until the LED shown illuminates.
Then using just the
up
or
doWn ArroW
adjust the End point value. The End point
value must be higher than the Plateau set
point.
note
Values set for all of the above will be
retained after the unit is powered off.
start switch Test program
Once the P, r , E values have been set the
Switch test can be started by pressing and
holding the Test button for a few seconds.
The display will show a flashing P and the
current block temperature which will flash
until the Plateau temperature is reached.
When the Plateau temperature has
stabilised the P will display continuously.
Then use the
up ArroW
button to
proceed to the next temperature with the
controlled heating ramp rate.
set
Test
E2000
Adjust (End point)
set
Test
P 1500
run sWITCh TEsT
set
Test
E2000
End point
sWITCh TEsT -End point
To check the End point for the final set
temperature of the switch test. Press the
Set button four times. (If checking after
adjusting the ramp rate press the Set
button once). End point values must be
higher than the Plateau set point values
and can be set to 700°C.