![Bibby Sterilin Techne TECAL 700 Operator'S Manual Download Page 21](http://html.mh-extra.com/html/bibby-sterilin/techne-tecal-700/techne-tecal-700_operators-manual_2749053021.webp)
TECAL 700 spECIfICATIon
Temperature
Temperature range (min set point 10°C or 50°F)
Amb.+25°C to 700°C (or 1292°F)
Temperature set point precision
0.01
Display Accuracy at 660°C
±0.25°C
Display Accuracy at 425°C
±0.2°C
Display Accuracy at 100°C
±0.1°C
Set point Display
°C or °F
Switch test ramp rate
0.1 to 10.0°C or 0.1 to 18.0°F
stability
(in amb of 25°C) after 10 mins at temperature over 30 minute period
425 - 700°C
±0.10°C
100 - 425°C
±0.04°C
50 - 100°C
±0.02°C
Axial uniformity
(between 0-40mm from base of well)
425 - 700°C
±1.5°C
100 - 425°C
±1.5°C
50 - 100°C
±0.5°C
radial uniformity
425 - 700°C
±0.5°C
100 - 425°C
±0.1°C
50 - 100°C
±0.02°C
Loading effect
425 - 700°C
±0.04°C
50 - 425°C
±0.02°C
hysteresis
±0.07°C
heating/Cooling rate
Heating rate between 50°C and 700°C
25 minutes
Cooling rate between 700°C and 100°C
30 minutes
Cooling rate between 700°C and 50°C
30 minutes
19