21
8.1.3 Viewing Videos
When a video record is accessed the Video Replay screen is displayed.
Block
temperature
at specific
time
Selected
capillary
Video replay
window
Video
controls
Displays report
information
summary
Tagged event
selection
Capillary
selection
buttons
Figure 8: Video Replay Screen
The Video Replay screen is designed to allow the user to review the recorded video of a melt
determination and reprocess the results if required.
Block temperature at specific time:
The temperature of the sample block at the specific timeframe
displayed in the video replay window.
Report information:
Pressing the “Touch for automatically calculated values” button displays the
screen below. Each capillary tube has its start temperature, end temperature and clear point displayed.
In addition, the mean value of the start and end temperatures is calculated and displayed.
Press the “Hide” button to return to the Video Replay screen.
Tagged event selection:
During the melt determination additional events may occur that are
required to be recorded in the report. Pressing this button accesses the list of available events that can
be added to the analysis report.
Capillary selection:
In order to record a manual event you must first select the number of the
capillary tube it applies to by using these buttons.
Video replay window:
The Video replay window displays the images that were recorded during the
substances melt determination. To view the image in full screen mode press the video replay window.
Pressing the screen again will return the device to the normal Video Replay screen.
Video controls:
The Video controls are used to control the playback of the substances melt
determination video.
Summary of Contents for Stuart SMP40