8 The option of saving this new method as a
new preset method will be given if the
sample name has been changed. Select “Yes”
to save the method for future use or select
“No” to proceed without saving the method.
9 If changes are made to the temperature
settings the user will be asked whether they
wish to overwrite the existing preset method
of that name. Select “Yes” to save the new
method settings for future use or select “No”
to proceed without saving the new settings.
13
6.3 Carry out a Rapid-Melt
If the user is unsure of the start temperature that should be specified in a melt determination method
they can use the rapid melt function to quickly determine an approximate melting temperature.
In the Measurement menu screen touch the “Carry out a rapid-melt” section of the screen.
1 The rapid melt function will heat the sample
from 50.0˚C at 20.0˚C/min until the sample
has melted.
2 The results will be reported on the screen as
shown in the example opposite.
3 The results from the rapid melt analysis can
then be used to facilitate the setting of the
start temperature on a more accurate melt
determination.