15
Check for loose connections on the heat sensor (pyrometer) wiring.
While the heater is running at temperature, gently wiggle the pyrometer cable and have
someone watch for sudden temperature changes. If this happens – replace the cable.
Problem
Positioning Beam does not work
Possible Cause and Corrective Action
Check that the power supply board is supplying approximately 3.25 vdc (TB 3). Check that the wires are secure
on the power supply board and on the push button. Open the cover on the heat sink assembly on the back of
the heater and retest for the 3.25 v. If the power supply is functioning and there is no voltage at the heat sink
assembly open the Raytek pyrometer and test for the voltage (the 2 wire nuts). – replace the Positioning
Beam.
Temperature Controller Error Codes
Display
Problem
Actions
Err.H
Open Sensor
Check sensor, wiring,
and input
Err.L
Reversed Sensor
Check Sensor polarity
LPbr
Loop Break
Correct problem and
Reset controller
0100
Checksum Error
Press Any Key to
perform a soft reset
and reinitialize
controller
0101
RAM Error
0202
Defaults Loaded
0303
EEPROM Write Failure
3865
Power Fail Resume
Feature Disabled
No further resume
actions available
36 Plus other 2 digit code
Unexpected or invalid
interrupt
Reset Controller
Ramp/ Soak Errors
02= Recipe Empty (i.e. no non-zero ramp times)
05= Insufficient Setpoint- Process Value Deviation
Summary of Contents for AutoCure 5000
Page 1: ...January 2016 AutoCure 5000 Portable Repair Arm Assembly and Operating Manual AC5 2116 380...
Page 18: ...18 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 19: ...19 Mechanical Drawings and Parts Lists Pyrometer...
Page 20: ...20 CAUTION...
Page 22: ...22...
Page 24: ...24...
Page 26: ...26 Heater Head Assembly ACH5 2116...
Page 40: ...40 SECTION J J SCALE 1 8 K DETAIL K SCALE 1 4 88 53 89 64 63 83 86 16 17...