5
4. Product Diagram
Figure 1
1
2
4
3
7
6
Figure 1:
1. Tabletop
2. Suction System (C1/C3 model only)
3. Water bottle (2)
4. Height Adjustment Knob
5. Air Compressor (C1/C2 model only)
6. Foot Pedal Control
7. 2.5-inch Castors
Page 1: ...Comfortek Dental Cart System Instruction for Use ...
Page 2: ...tions 4 4 Product Diagram 5 5 Symbols Legend 7 6 Unpacking Set Up 8 7 Operating Instructions 10 8 Daily Cleaning and Maintenance 13 9 Troubleshooting Guide 16 10 Replacement Parts 18 11 Water air Connection Diagram 19 12 Electrical Connection Diagram 20 13 Technical Support 21 14 Warranty 21 ...
Page 3: ...in use please turn the unit off immediately and contact Beyes or an authorized dealer Do not dismantle the machine This product should be only be repaired by an authorized Beyes technician The user of the product shall be responsible for any failure caused by improper use faulty maintenance or improper repair and damage caused by a technician that was not authorized by Beyes 2 Product Introduction...
Page 4: ...mum continuous operation time is 15mins Tank Capacity 5L 1 32gal Air pressure 7 5bar 108psi 2 1 Application Scope This equipment is a mobile cart system for mobile dentistry Comfortek Dental Cart System is available in 4 models Dental Cart System Air Compressor Suction system C1 C2 C3 C4 2 2 Product size Length 45 1cm 17 75 Width 45 1cm 17 75 Min height 86 2cm 33 94 Max height 115cm 45 28 Weight 5...
Page 5: ...roduct Diagram Figure 1 1 2 4 3 5 7 6 Figure 1 1 Tabletop 2 Suction System C1 C3 model only 3 Water bottle 2 4 Height Adjustment Knob 5 Air Compressor C1 C2 model only 6 Foot Pedal Control 7 2 5 inch Castors ...
Page 6: ...oggle Switch 7 Safety Toggle Switch to depressurize water system 8 Pressure Gauge for Handpiece HS LS 9 HDPE bottle 1000ml Optional 10 Chip Air Adjustment Handpiece 11 Drive Air Adjustment Handpiece 12 Water Flow Adjustment Knobs Handpiece 13 Water Flow Adjustment Knob Piezo Scaler 14 Power setting Frequency Control Piezo Scaler 15 Bottle 1 16 Bottle 2 1 2 3 4 5 8 6 7 9 12 13 14 15 16 10 11 ...
Page 7: ... 5 Air Regulator Valve 6 Main Power Switch 7 Overcurrent protector 1 2 3 4 5 6 7 5 Symbols Legend Symbol Symbol Switch on Handpiece water regulator Switch off Handpiece chip air regulator Warning Handpiece drive air regulator Foot control Water cooling Water bottle toggle switch Safety Toggle Switch to depressurize water system ...
Page 8: ...uence from left to right 1 Air water syringe 2 Suction install saliva injector tip or suction curved tip 3 High speed Low Speed Handpiece 4 High speed Low Speed Handpiece 5 To connect the piezoelectric scaler handpiece to the cord align the pins on the handpiece with the cord coupler and gently but firmly push the handpiece into the cord coupler Do not twist turn or unscrew the handpiece while ins...
Page 9: ...e between 80 psi 5 5 bar and 108 psi 7 5 bar The compressor will automatically stop when the pressure reaches 108 psi 7 5 bar Figure 8 Figure 9 Figure 10 Figure 8 Connect the power plug of the power and press the main power switch when the low power supply voltage and load startup may trigger overload protector to cut off the motor s power At this time the device needs to stand for approximately 1...
Page 10: ...r bottle safety toggle switch A is the water bottle toggle switch the left side is to set water source to 1 water bottle 1 the right side is to set water source to 2 water bottle 2 B is the safety toggle switch to pressurize or depressurize the water system Turn the knob all the way to the left to pressurize the water system and all the way to the right to depressurize the water system Warning Mak...
Page 11: ...11 7 4 Air Water syringe Check 7 5 Drive air pressure and chip air regulator Water flow adjustment knobs Handpiece chip air regulator Handpiece drive air regulator ...
Page 12: ...h speed handpiece 0 2 0 3 MPa 30 40 PSI Low speed handpiece A B Please select and install the suction tip that best meets your treatment needs A The curved suction tip is installed by directly inserting into the suction handle B The saliva injector tip is installed by directly inserting into the suction handle ...
Page 13: ...disinfectant Use a solution that does not contain chloride or iodine to prevent staining of the unit 8 2 The suction hose is exposed to pathogenic secretions saliva and blood on a daily use Therefore for hygiene reasons the used suction hose must be cleaned after each operation especially if bleeding occurs during the operation If treatment is prolonged the suction hose must be cleaned at least on...
Page 14: ... Remove the High Speed Handpiece and Low Speed Handpiece from the holder 2 Disconnect the handpiece from the tubing cord 3 Adjust the water regulator for all equipment to maximum flow 4 Put the treatment cords over a sink or a bucket 5 Step on the foot pedal for 20 to 120 seconds to wash the waterline of the treatment instrument 6 Put the handpieces back on the holder Foot Pedal Improve its stabil...
Page 15: ...y specific directives standards and guidelines for sterilization temperature and times Figure 12 8 4 Piezoelectric Scaler and Handpiece Daily Cleaning Disconnect the cord and Piezoelectric handpiece and clean handpiece with a disinfectant non chloride non iodine Do not immerse in any solution including ultrasonic or cold sterile solutions as this will damage the handpiece and void Warranty Make su...
Page 16: ...nce a day Please check the level of the dirt cup and empty it if necessary Figure 15 Figure 13 Figure 14 Figure 15 9 Troubleshooting Guide 9 1 Error No Power 1 Ensure the power cord is plugged into a live wall circuit of the appropriate voltage 2 The machine end is firmly plugged into machine 3 The main switch is on 4 Ensure the overcurrent protector hasn t popped 9 2 Error The tank will not press...
Page 17: ...l to be used 2 Adjust pressure regulator on underside of tray top the front most dial controls the chip air while the rear controls the tool drive air 3 For the scaler make sure that the control knob for the strength of the vibrations is turned to the correct amount 9 6 Error No suction 1 Make sure you are fully depressing the small push button switch on the foot pedal 2 Ensure the suction control...
Page 18: ... 4 1 Straight nose attachment 19054 DPA FIRM 19053 DPA SOFT UL2420 Tip Wrench UL2022 Scaler handpiece UL2300 E G1 UL2301 E G2 UL2302 E G4 UL2305 E P1 UL2306 E P3 C2360 Turbine HSP3008 Precision coupler o ring set 804030203 4 hole Gasket 871032227 Water Bottle 1 2 871031876 Suction HDPE bottle 1000ml ...
Page 19: ...19 11 Water air Connection Diagram ...
Page 20: ...20 12 Electrical Connection Diagram ...
Page 21: ...ponents or products may be used and or refurbished components or products provided they are of like quality and specifications as new components or products 14 2 Applicable Waranty Period The applicable warranty period measured from the date of invoice to the original user shall be as follows Comfortek Dental Cart Systems are warranted for a period of 36 months Air Driven High Speed Handpieces are...
Page 22: ...s made by any person or entity other than BEYES 8 matching of color grain or texture except to commercially acceptable standards 9 changes in color caused by natural or artificial light 10 custom manufactured products 11 alterations or modifications to the product by any person or entity other than BEYES 12 Products that would otherwise by covered under Sections 1 and 2 of this limited warranty bu...
Page 23: ...ties of merchantability or fitness for a particular purpose This warranty is limited to the repair or replacement of defective parts 14 6 Statue of Limitations No actions may be brought against BEYES for breach of this limited warranty or implied warranty if any or for any other claims arising out of or relating to the products more than ninety 90 days following expiration of the limited warranty ...
Page 24: ...naplein 10 Ie Verd 2595AA The Hague Netherlands Tel 31645171879 English 31626669008 Dutch Email info beyes ca Website www beyes ca Printed in Canada ENI047 Rev 1 29 10 21 Federal law restricts this device to sale by or on the order of a dentist physician or any other practitioner licensed by the law of the states in which he or she practices to use or order the use of this device 0197 ...