8.3 High Speed and Low Speed Handpieces Lubrication
Oil High Speed and Low Speed handpieces prior to first use, and oil after every
use.
1. After cleaning the handpiece, apply lubricant before sterilization. lubricant is
placed in the smaller of the 2 large holes at the base of the handpiece. Use only a
high quality lubricant such as BeyesLube.(Figure 11)
2. Flush through with air for 30-45 seconds and wipe any excess lubricant from
handpiece.
3. Place the handpiece in a sterilization pouch and follow your country specific
directives, standards, and guidelines for sterilization temperature and times.
(Figure 12)
8.4 Piezoelectric Scaler and Handpiece Daily Cleaning
Disconnect the cord and Piezoelectric handpiece and clean handpiece with a
disinfectant (non-chloride, non-iodine). Do not immerse in any solution, including
ultrasonic or cold sterile solutions, as this will damage the handpiece and void
Warranty. Make sure to remove any tips from the handpiece prior to sterilization.
1. Dry handpiece with a clean cloth
2. Place handpiece in sterilization bag and sterilize for 3 minutes at a maximum of
135 °C (Figure 12)
*Additional handpieces and tips can be purchased to accommodate sterilization turn
around time.