2. Function introduction

2.1 Interface display introduction

2.1.1 Icon introduction

Picture 4 (b)

Picture 4 (a)

Mute

Woof

Alt

Bluetooth open when charging

Mediant

Bluetooth is starting or is being initialized

Bluetooth normal open

Bluetooth has been built to connect

Battery power status indicator: 0%, 25%, 50%, 75%, 100%

Battery charging status display

Demo mode, this icon will appear when the device is in demo mode

Connection test, when the measurement line short circuit this icon will
This icon will appear when the measurement results are in pre - setting the root tip
This icon will appear when the measurement result is beyond the root tip

Summary of Contents for AL2040

Page 1: ...Apex Locator ApexPilot Instruction for Use ...

Page 2: ...ssification of Device Classification by type of protection against electric shock class II devices Classification by degree of protection against electric shock applied part type B Degree of protection against ingress of water IPX0 Sterilization or disinfection method refer to Sterilization part in the manual Degree of safety of application in the presence of a flammable anesthetic mixture with ai...

Page 3: ...Bluetooth Function 2 4 Gyroscope 2 5 Charging and USB introduction 3 Installation P10 11 3 1 Connect the measurement line 3 2 Demo function 3 3 Battery charging 4 Product operation P11 15 4 1 Operation Notice 4 2 Usage requirement 4 3 Instruction 5 Trouble shooting P16 18 6 Cleaning and disinfection P18 7 Storage maintenance and transportation P19 8 Environment protection P19 9 After service P20 1...

Page 4: ...c calibrating ensures the measurements are accurate Equipped with high precision gyroscope switch the display mode according the state of display Bluetooth wireless transmission get rid of the long line File clip lip hook and touch probe can be autoclaved under high temperature and high pressure avoiding cross infection effectively Built in non removable battery Angle adjustable easy to adjust the...

Page 5: ...r Implantable electronic device and patients who are advised not to use the electric equipment like electric shaver electric blower do not recommend the use this product 1 6 Main technical specifications 1 6 1 Battery 3 7V 1000mAh 1 6 2 Adapter 100 220V 50 60Hz 0 25A 1 6 3 Consumption power 0 6W 1 6 4 Screen 3 5 TFT Wide angle of view LCD 1 6 5 Buzzer alert the buzzer will alert when the endo file...

Page 6: ... normal open Bluetooth has been built to connect Battery power status indicator 0 25 50 75 100 Battery charging status display Demo mode this icon will appear when the device is in demo mode Connection test when the measurement line short circuit this icon will This icon will appear when the measurement results are in pre setting the root tip This icon will appear when the measurement result is be...

Page 7: ... slowly close to the root tip color scale will gradually fill the alarm sound will be more rapid non silence state When the needle reached the preset of the root file the alarm will ring often to remind that has reached the apex When file tip needle penetration red dot prompt is displayed below the enlarged area with accompanying alarm Picture 5 c is the position of the apical region of the root T...

Page 8: ...Bluetooth Function Bluetooth of this product can establish a wireless signal transmission with our endo motor In the connection state the endo motor can send the measured data by the Bluetooth signal to the apex locator to display 2 3 1 Bluetooth link for the first time Device 1 pc apex locator ApexPilot 1 pc endo motor ApexPilot All in one Environment make sure no other Bluetooth signal interfere...

Page 9: ...irection 2 5 Charging and USB introduction Using professional lithium battery charging management chip charging current can be up to 600mA There is a slight fever during charging which is normal Please use the original charger to charge the battery we will not assume any responsibility for the damage caused by using other chargers The USB micro interface is not only the measurement line interface ...

Page 10: ...Picture 7 Picture 8 a Picture 8 b Picture 9 3 1 3 Test the wire connecting before use Turn on the unit not demo mode confirm that Bluetooth connection is not established Make sure the measuring line inserting into the USB interface and well connected with the file clip and lip hook Make the lip hook touch the file clip Pic 9 when this signal appears on the screen and stable the connection is norma...

Page 11: ... charger and charging line plug into the USB interface for charging 3 3 3 If in Bluetooth mode or demo mode when charging the display icon of the battery power will become green with charging animation effect if the battery is full the power icon will keep full green 3 3 4 When charging in normal detection mode battery charging status display will appear in full screen when the battery is full it ...

Page 12: ...nd high pressure before use 4 1 8 Do not disassemble the product without permission once demolition there will no warranty 4 2 Usage requirement 4 2 1 To measure according to the specific description in the manual 4 2 2 The dentists should have the knowledge of teeth position and average length and the skill to operate the device 4 2 3 An fully exposed access cavity to show the pulpal cabin 4 2 4 ...

Page 13: ... will be on a conducting state which will cause an inaccurate result while measuring The measurement can be continued when the bleeding is stopped Pic 11 The tooth crown is broken The tissue of the gingival may reach the cavity of the endo hole at the broken point which will cause inaccuracy because of the electronic conduction The measurement can be continued when the crown is fixed by gypsum or ...

Page 14: ... t show the accurate position of the apical foramen Picture 16 4 3 Instruction 4 3 1 Please insert the plug into the socket on the right side of the main unit turn on the device The instrument is normally on and without warning which indicates that the instrument is normal Please refer to the troubleshooting section when there is a warning 4 3 2 Connect the file clip and lip hook Pic 9 this icon a...

Page 15: ...the measurement inaccurate Please reference the picture 20 a to grip the needle file if as picture 20 b it can t properly measure the length of the root canal due to the improper force the front of the root canal pin is easy to wear Picture 19 Picture 20 4 3 6 When the file reaches the apex adjust the rubber piece set on the endo file to the reference point incisal edge or fossa edge then pull out...

Page 16: ...rmally and display Calibration failure please press power to exit if not the device can be used normally Please note if in this situation there must be something wrong with the circuit inside the measurement will be not accurate please contact with your dealer to solve this problem Bluetooth is on but can t be connected Check if the Bluetooth display is normally and not connected yet For the first...

Page 17: ...t approaches the apex Is the tooth surface covered with cutting debris or chemical solutions Clean entire tooth surface Is the file touching the gingival tissue This will cause the canal length indicator bar to suddenly jump all the way to the OVER Is there pulp tissue left inside the root canal An accurate measurement cannot be obtained if a large amount of pulp tissue is left inside the root can...

Page 18: ...apical foramen Is root canal blocked Canal length indicator will return to normal when the file reaches apical constriction Is root canal extremely dry Moisten the root canal with hydrogen peroxide or a saline solution A small file in a large root canal Choose right size of the file 6 Cleaning and disinfection 6 1 Wipe the main unit and measuring line with a piece of gauze dampened with ethanol 6 ...

Page 19: ...horized person or authorized after service center 7 2 2 Keep the product in a dry storage condition 7 2 3 Do not throw beat or shock the product 7 2 4 Do not smear the product with pigments 7 3 Transportation 7 3 1 Excessive impact and shake should be prevented in transportation Lay it carefully and lightly don t invert it 7 3 2 Don t put it together with dangerous goods during transportation 7 3 ...

Page 20: ...this equipment free of charge if there s quality problem Please refer to the warranty card 0197 This conforms to CE European Directive of Medical equipment directive 93 42 EEC Follow the waste of electric and electronic equipment WEEE Directive Class II equipment Double insulated fulfilling legal requirement of IEC 60601 1 consult accompanying document user manual Type B particular protection agai...

Page 21: ...lectromagnetic Environment guidance RF Emissions CISPR 1 Group 1 The appliance use RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 1 Class B The ApexPilot is suitable for use in at I establishments including domestic establishments and those directly connected to the publ...

Page 22: ... hospital environment Surge IEC 61000 4 5 1 kV differential mode 1 kV differentia mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT For 0 5 cycles 40 UT 60 dip in UT for 5 cycles 5 UT 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 s 5 ...

Page 23: ... site survey shouId be less than the compliance level in each frequency range Radiated RF GB T17626 3 3V m 80MHz 2 5GHz 3V m NOTE UT is the ac mains voltage prior to application of the test level NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in at I situations Electromagnetic propagation is affected by absorption and reflection from structure...

Page 24: ...unications equipment Rated maximum output power of transmitter Watts W Separation distance according to frequency of transmitter in meters Meters m 150 KHz 80 MHz d 1 2 P 80 MHz 800 MHz d 1 2 P 800 MHz 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance...

Page 25: ... installation procedures for the cable connection Apex locator should not be used stacked or close to other equipment If adjacent or stacked use is necessary you should observe and verify it works normally Using the specified peripherals Avoid to use not specified equipment otherwise it may cause the lower performance of EMC Check the function of the corresponding sections of vital signs range thi...

Page 26: ...ranty period measured from the date of invoice to the original user shall be warranted for a period of 12 months 12 3 EXCLUSIONS This limited warranty does not cover and BEYES shall not be liable for the following 1 Defects damage or other conditions caused in whole or in part by misuse abuse negligence alteration accident freight damage negligent storage tampering or failure to seek and obtain re...

Page 27: ...lth or dental market as the case may be in which purchaser intends to use the product 12 4 EXCLUSIVE REMEDY CONSEQUENTIAL DAMAGES DISCLAIMER BEYES OBLIGATION UNDER THIS LIMITED WARRANTY IS THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS BEYES SHALL NOT BE LIABLE FOR AND HEREBY DISCLAIMS ANY DIRECT SPECIAL INDIRECT INCIDENTAL EXEMPLARY OR CONSEQUENTIAL DAMAGES OR DELAYS INCLUDING BUT NOT LIMITED TO DA...

Page 28: ... use of this device Beyes Dental Canada Inc is not responsible for any typographical errors All brands are holders of their respective trademarks 0197 Beyes Dental Canada Inc 23 595 Middlefield Road Toronto Ontario M1V 3S2 Canada Tel 1 855 603 1888 Fax 1 855 720 1228 Email info beyes ca Web www beyes ca Lotus NL B V Koningin Julianaplein 10 Ie Verd 2595AA The Hague Netherlands Tel 31645171879 Engl...