15
Notice
• When grip the root canal with a needle file, please grip the upper of the metal part
(near the root canal at the needle handle). If you grip the lower partb(blade or moving
part), it will wear the metal part of the file folder and the resin part. (Pic 19)
• When measuring the length of the root canal, please don't use the metal needle file. If
you operate the device without the dentistry glove, it will cause leakage and the result
of the measurement will be inaccurate. Therefore, please use the resin needle file
and remember don't touch the metal part with finger.
• Please don't use the worn file clip, it will make the result of the measurement
inaccurate.
• Please reference the picture 20 (a) to grip the needle file, if as picture 20 (b) it can't
properly measure the length of the root canal due to the improper force, the front of
the root canal pin is easy to wear.
Picture 19
Picture 20
4.3.6 When the file reaches the apex, adjust the rubber piece set on the endo file to
the reference point(incisal edge or fossa edge), then pull out the endo file, measure
the length between the top of the file and the rubber piece, and this is the working
length of the tooth.
4.3.7 The components that touch body should be autoclaved under high
temperature and high pressure. The shell and measuring line should be cleaned by
75% alcohol.
4.3.8 Press the power key more than 1 second to turn off the device, the device will
automatically shut off after 10 minutes without operation.