9

Disinfection

Please carefully observe the points below when disin-

fecting your atomiser and accessories. We recommend 

disinfecting the individual parts on a daily basis after the 

last usage as a minimum measure.

(All you need for this is a little white vinegar and distilled 

water.)

• First, clean the atomiser and accessories as described 

in the “Cleaning” section.

• Place the disassembled atomiser, mouthpiece and 

nosepiece in boiling water for 5 minutes.

• For the remaining accessories, use a vinegar solution 

consisting of ¼ vinegar and ¾ distilled water. Make 

sure that the volume is sufficient to fully submerse the 

parts, such as the atomiser, mask and mouthpiece, in 

the solution.

• Leave the parts in the vinegar solution for 30 minutes.

• Rinse the parts with water and dry them carefully with 

a soft cloth.

 Important

Do not boil or autoclave the compressed air hose or the 

masks.

• Put the parts together again when they are completely 

dry and place them in a dry, sealed container.

 Note

• Please ensure that the parts are completely dried af-

ter cleaning, otherwise the risk of bacterial growth is 

increased.

• Use cold disinfection solution in accordance with the 

manufacturer’s instructions.

Drying

• Place the individual parts on a clean, dry and absor-

bent surface and leave them to dry completely (at least 

4 hours).

Durability of materials

• As with any plastic parts, atomisers and their accesso-

ries are affected by a certain amount of wear and tear 

when used and hygienically prepared on a frequent ba-

sis. Over time, this can lead to a change in the aerosol, 

which can have a negative effect on the efficiency of the 

treatment. We therefore recommend that you replace 

the atomiser and other accessories after a year.

•  The following point should be observed when selecting 

the cleaning product or disinfectant: only use a mild 

cleaning product or disinfectant in the concentration 

and volume prescribed by the manufacturer.

Storage

• Do not store in damp conditions (such as in a bathroom) 

and do not transport with any damp items.

•  When storing and transporting, protect from prolonged 

direct sunlight.

• The accessories can be stowed securely in the ac-

cessory compartment. Store the device in a dry place, 

ideally in the original packaging.

9. Disposal

For environmental reasons, do not dispose of the 

device in the household waste.

Please dispose of the device in accordance with 

EC Directive – 

WEEE

 (Waste Electrical and Electronic 

Equipment). If you have any questions, please contact 

the local authorities responsible for waste disposal.

10. Troubleshooting

Problem/

question

Possible cause/remedy 

The atomiser 

produces no 

or too little 

aerosol.

1.  Too much or too little medication 

in the atomiser. 

Minimum: 2 ml,  

Maximum: 6 ml.

2.  Check nozzle for blockages. 

Clean nozzle if necessary (e.g. by 

rinsing out). Then start using the 

atomiser again.

    IMPORTANT:

 Carefully pierce the 

fine holes from the underside of the 

nozzle only.

3.  Atomiser not held vertically.

Summary of Contents for IH 58

Page 1: ...IH 58 EN Nebuliser Instructions for use...

Page 2: ...recovery Ask your doctor or pharmacist for fur ther information about the potential applications The device is suitable for domestic use Inhaler medica tion should only be used following instructions...

Page 3: ...e dosage and the frequency and duration of inhalation Only use medication prescribed or recommended by your doctor or pharmacist Please note For treatment only use parts indicated by your doctor accor...

Page 4: ...We therefore recommend that you obtain a replacement device or medication the latter should be agreed with your doctor Should you require an adapter or extension lead this must meet the applicable saf...

Page 5: ...p with filter 5 Hose connector Overview of atomiser and accessories 6 Compressed air hose 7 Atomiser 8 Medication container 9 Atomiser insert 10 Mouthpiece 11 Adult mask 12 Children s mask 13 Nosepiec...

Page 6: ...twisting the top anticlock wise against the medication container 8 Place the atomiser insert 9 into the medication container 8 Ensure that the cone for administering medication fits well on the cone f...

Page 7: ...sensitive bronchial system medications containing essential oils may under certain conditions cause an acute bronchos pasm a sudden cramp like restriction of the bronchi with shortness of breath Consu...

Page 8: ...container 8 Reassembly is carried out in reverse order Cleaning Important The device must be switched off disconnected from the mains and allowed to cool down each time before cleaning The atomiser an...

Page 9: ...rs and their accesso ries are affected by a certain amount of wear and tear when used and hygienically prepared on a frequent ba sis Over time this can lead to a change in the aerosol which can have a...

Page 10: ...hould not be left alone Inhalation us ing the mask takes longer This is due to technical reasons You breathe less medication per breath through the holes of the mask than via the mouthpiece The aeroso...

Page 11: ...guaranteed We reserve the right to make technical changes to improve and develop the product This device and its accesso ries comply with the European standards EN60601 1 EN60601 1 2 CISPR 11 IEC61000...

Page 12: ...r local retailer in the first in stance see the attached International Service list of service addresses The buyer will then receive further information about the processing of the warranty claim e g...

Reviews: