If you are using the adult mask [11], you can secure it 

to your head with the adult mask strap. There is also a 

strap for the children’s mask [12].

When using a mask to inhale, take care to ensure the 

mask fits well and the eyes are unobstructed.

• Before the treatment, pull the atomiser upwards out 

of the holder [3].

•  Start the nebuliser using the On/Off switch [1].

• Spray mist pouring out of the atomiser indicates that 

the device is operating correctly.

6.5 Treatment

• When inhaling, sit upright and relaxed at a table and 

not in an armchair to avoid compressing the airways 

and therefore impairing the effectiveness of the treat-

ment.

• Breathe in the atomised medication deeply.

 Important

The device is not suitable for continuous operation; after 

30 minutes of operation it must be switched off for 30 

minutes.

 Note

During the treatment, hold the atomiser straight (vertical-

ly), otherwise the atomisation will not work and faultless 

functionality is not guaranteed.

Depending on the battery charge status and accessory 

used, the inhalation time stated may vary.

 Important

Essential oils, cough syrups, gargling solutions and drops 

to be used as a rub or in a steam bath are wholly unsuit-

able for inhalation using a nebuliser. These additions are 

often viscous and can impair the correct functioning of 

the device and therefore the effectiveness of the applica-

tion in the long-term. For individuals with a hypersensitive 

bronchial system, medications containing essential oils 

may under certain conditions cause an acute bronchos-

pasm (a sudden cramp-like restriction of the bronchi with 

shortness of breath). Consult your doctor or pharmacist 

in relation to this matter.

6.6 Stopping inhalation

Once the mist is only coming out in an irregular flow or 

if the sound changes when inhaling, you can stop the 

treatment.

• Switch off the nebuliser after treatment using the On/

Off switch [1] and disconnect it from the mains.

• Place the atomiser [7] back in its holder [3] after the 

treatment.

6.7 Cleaning

See “Cleaning and disinfection” on page 8.

7. Changing the filter

In normal operating conditions, the air filter must be re-

placed after approx. 100 operating hours or one year. 

Please check the air filter regularly (after 10–12 nebu-

lisation procedures). Replace the used filter if it is very 

dirty or clogged. If the filter has become damp, it must 

also be exchanged for a new filter.

 Important

• Do not attempt to clean the used filter and reuse it.

• Only use the manufacturer’s original filter, otherwise 

your nebuliser may become damaged and sufficiently 

effective treatment cannot be guaranteed.

• Do not repair or maintain the air filter while it is in use.

• Never operate the device without a filter.

To replace the filter, proceed as follows:

 Important

• First switch the device off and disconnect it from the 

mains.

• Allow the device to cool down.

1. Pull off the filter cap [4] 

towards the front.

 Note

If the filter remains in the device after the cap has been 

removed, take the filter out of the device, e.g. with twee-

zers or similar.

4

Summary of Contents for IH 58

Page 1: ...IH 58 EN Nebuliser Instructions for use...

Page 2: ...recovery Ask your doctor or pharmacist for fur ther information about the potential applications The device is suitable for domestic use Inhaler medica tion should only be used following instructions...

Page 3: ...e dosage and the frequency and duration of inhalation Only use medication prescribed or recommended by your doctor or pharmacist Please note For treatment only use parts indicated by your doctor accor...

Page 4: ...We therefore recommend that you obtain a replacement device or medication the latter should be agreed with your doctor Should you require an adapter or extension lead this must meet the applicable saf...

Page 5: ...p with filter 5 Hose connector Overview of atomiser and accessories 6 Compressed air hose 7 Atomiser 8 Medication container 9 Atomiser insert 10 Mouthpiece 11 Adult mask 12 Children s mask 13 Nosepiec...

Page 6: ...twisting the top anticlock wise against the medication container 8 Place the atomiser insert 9 into the medication container 8 Ensure that the cone for administering medication fits well on the cone f...

Page 7: ...sensitive bronchial system medications containing essential oils may under certain conditions cause an acute bronchos pasm a sudden cramp like restriction of the bronchi with shortness of breath Consu...

Page 8: ...container 8 Reassembly is carried out in reverse order Cleaning Important The device must be switched off disconnected from the mains and allowed to cool down each time before cleaning The atomiser an...

Page 9: ...rs and their accesso ries are affected by a certain amount of wear and tear when used and hygienically prepared on a frequent ba sis Over time this can lead to a change in the aerosol which can have a...

Page 10: ...hould not be left alone Inhalation us ing the mask takes longer This is due to technical reasons You breathe less medication per breath through the holes of the mask than via the mouthpiece The aeroso...

Page 11: ...guaranteed We reserve the right to make technical changes to improve and develop the product This device and its accesso ries comply with the European standards EN60601 1 EN60601 1 2 CISPR 11 IEC61000...

Page 12: ...r local retailer in the first in stance see the attached International Service list of service addresses The buyer will then receive further information about the processing of the warranty claim e g...

Reviews: