Beurer GL44 lean
33
9 T
ECHNICAL
SPECIFICATIONS
Type
GL 44
Dimensions (W x H x D)
52 x 95 x 16 mm
Weight
44 g incl. batteries
Power supply
2 x 3 V CR 2032 button cell batteries
Battery life
1000 measurements
Measured value memory
480 measured values with date/time
Data retained when batteries are changed
Average values
for 7, 14, 30, 90 days
Automatic switch-o
ff
2 minutes after last actuation
Storage/transport temperature
Temperature: +2°C – +30°C
Relative humidity: < 90%
Operating ranges
Temperature: +10°C – +40°C
Relative humidity: < 90% non-condensing
Measuring range, glucose
Glucose: 20 – 630
mg/dL (1.1 - 35.0
mmol/L)
Blood sample
Capillary whole blood
Amount of blood
0.6 microlitres
Blood glucose measurement
duration
Approx. 5 seconds
Calibration
Plasma
Testing method
Amperometric bio sensor
Use
Suitable for self-testing
System function test
Each time device is switched on
The serial number is located on the device or in the battery compartment.
EMC
This device complies with the European standard EN 61326 and is subject to specific precautions
with regard to electromagnetic compatibility. Please note that portable and mobile HF communica-
tion systems may interfere with this device. For more details, please contact our Customer Services
at the address indicated.
Test strip functionality
Test strips enable a quantitative measurement of the glucose level in fresh whole blood. When the
gap for taking blood comes into contact with a drop of blood, it is automatically filled by simple cap-
illary action. The blood is sucked into the absorbing gap on the test strip and the monitor measures
the blood glucose level in the blood.
The test is based on the measurement of an electric current that is generated by the chemical reaction
of the glucose with the enzyme glucose dehydrogenase (Aspergillus oryzae) on the strip.