9. Error messages/troubleshooting
In the event of errors, the error message
EE
appears on the
display.
Error messages may appear if:
• Systolic or diastolic pressure could not be measured (
EE1
)
• The cuff is faulty or fastened too tightly or too loosely (
EE2
)
• The pump pressure is higher than 290 mmHg (
EE8
)
• The data could not be sent via
Bluetooth
®
(
EE7
)
10. Technical specifications
Model
SR BM 1
Type
BM 57
Measurement me-
thod
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Measurement range
Cuff pressure 0 – 300 mmHg,
Systolic 50 – 250 mmHg,
Diastolic 40 – 180 mmHg,
Pulse 40 – 180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard deviation
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 94 mm x W 161 mm x H 64 mm
Weight
317 g (without cuff)
Cuff size
23 to 43 cm
Permissible operating
conditions
10°C to +40°C, ≤85% relative humidity
(non-condensing)
Permissible storage
and transport con-
ditions
-20 to + 55°C, ≤85% relative humidity,
800 – 1050 hPa ambient pressure
Power supply
2x 1.5 V
AA batteries
Battery life
Lasts for approx. 150 measurements
depending on the blood pressure and
pump pressure as well as the number
of
Bluetooth
®
connections.
Accessories
Cuff, instructions for use, storage
pouch
Classification
Internal supply, IP22, no AP or APG,
continuous operation, application part
type BF
Data transfer via
Bluetooth
®
wireless
technology
The blood pressure monitor uses
Blue-
tooth
®
low energy technology,
2402 MHz – 2480 MHz frequency band,
transmission power max. 0 dBM
Compatible with
Bluetooth
4.0 smart-
phones/tablets
List of supported smartphones/tablets