141
Table 5:
Guidance and manufacturer’s declaration – electromagnetic immunity
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the
Device should assure that it is used in such an environment
Radiated RF
IEC61000-4-3
(Test speci-
fications for
ENCLOSURE
PORT IMMU-
NITY to
RF wireless
communica-
tions equip-
ment)
Test
Frequen-
cy
(MHz)
Band a)
(MHz)
Service a)
Modulation b)
Modulation b)
(W)
Distance
(m)
IMMUNI-
TY
TEST
LEVEL
(V/m)
385
380 –390
TETRA 400
Pulse
modulation b)
18 Hz
1,8
0,3
27
450
380 –390
GMRS 460,
FM c)
± 5 kHz de-
viation
1 kHz sine
2
0,3
28
710
704 – 787
LTE Band
13,
17
Pulse
modulation b)
217 Hz
0,2
0,3
9
745
780
810
800 – 960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
2
0,3
28
870
930
1720
1 700 –
1 990
GSM 1800;
CDMA
1900;
GSM 1900;
DECT;
LTE Band
1, 3,
4, 25; UMTS
Pulse
modulation b)
217 Hz
2
0,3
28
1845
1970
2450
2 400 –
2 570
Bluetooth,
WLAN,
802.11
b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz
2
0,3
28
5240
5 100 –
5 800
WLAN
802.11
a/n
Pulse
modulation b)
217 Hz
0,2
0,3
9
5500
5785
NOTE
U
T
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna
and the
ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
Summary of Contents for 60842
Page 3: ...3 2 1 3 4 6 5...
Page 69: ...69 2 3 3 3...
Page 70: ...70 3 3 3...
Page 72: ...72 6 3 25 45 7 1 2 5 2 3 TL 85 0 5 1 4 1 30 30 1 60 2 90 3 120 4 5...
Page 73: ...73 4 25 45 7 06 00 20 00 2 5 25 45 10 000 25 0 5 5000 30 1 2500 45 2 6 7 7 5 2...
Page 74: ...74 8 9 www beurer com 10 5 11 WEEE Waste Electrical and Electronic Equipment...
Page 77: ...77 LXCP62 II 240 100 240 50 60 1 5 A 24 2 5 IP24 II 13 MDR EU 2017 745...