TECHNICAL SERVICE BULLETIN
PB2 Bypass for Using External Nephros SSU Filter
TSB#
TSB2012008
Date
10/15/12
Page 1 of 1
OVERVIEW:
The PB2 (portable RO) comes standard with an internal .05 micron Pyrogen filter. Better
Water has created a simple solution for those customers who have expressed an interest
in using an external Nephros SSU Pyrogen capsule filter. This solution involves the
removal of the standard internal .05 micron Pyrogen filter and replacing it with a bypass
tube with quick-connect ends.
This bypass tube, part#
EQSUBPB2FT
is available from Better Water.
USE of the NEPHROS SSU CAPSULE FILTER:
When using the Nephros SSU capsule filter, it should be installed on the product hose between the PB2 and
the dialysis machine. A sample port should also be installed after this filter to draw water to mix bicarb and
for bacteria sampling,
REPLACEMENT INSTRUCTIONS:
1. Turn the power to the RO unit OFF.
2. Turn the water supply to the RO unit OFF.
3. Open the lab-cock, located inside the RO cabinet, and relieve filter-in and filter-out pressure. It is
advisable to use a small piece of tubing on the nipple of the lab-cock to route the draining water out of the
cabinet.
4. Close the lab-cock once the drain is complete.
5. Remove the .05 micron filter (
refer to the PB2 Operator Manual for instructions on how to remove the
filter
)
6. Attach the false/bypass tube in place of the filter. The connections will only allow it to be installed one
way.
7. Turn the water supply to the RO unit ON.
8. Open the lab-cock and allow water to flow until all air is purged. It is advisable to use a small piece of
tubing on the nipple of the lab-cock to route the draining water out of the cabinet.
9. Close the lab-cock once all the air is purged.
10. Install the Nephros SSU capsule filter on the product hose coming from the PB2.
11. Install a sample port after the Nephros SSU capsule filter.
12. The unit is now ready to run.
PB2 .05 False Tube
part# EQSUBPB2FT
Nephros SSU Pyrogen
Capsule Filter
NOTE
Better Water LLC recommends that one of these two filters must
be used with the PB2 to ensure water quality.
Summary of Contents for EQROPB01600
Page 1: ...REA 2380 PB2 Mediport Portable RO Operator Manual Better Water LLC rev Jul 2019 ...
Page 20: ...Better Water LLC rev Jul 2019 Page 16 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 36: ...Better Water LLC rev Jul 2019 Page 32 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 56: ...Better Water LLC rev Jul 2019 Page 52 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 80: ...Better Water LLC rev Jul 2019 Page 76 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 82: ...Better Water LLC rev Jul 2019 Page 78 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 84: ...Better Water LLC rev Jul 2019 Page 80 of 80 PB2 Mediport Portable RO Operator Manual ...