23
OPERATION MANUAL AND INSTRUCTIONS
EN
LIMITS
ISO 6789-1:2017
The Dynatester automatically calculates the limits to ISO 6789-1:2017 by the following method:
Deviation = (Displayed reading - Target value ) x 100
Target value
If the 100% target value is set above 10 Nm, the deviation must be within ± 4%.
For a 100% target value of 10 Nm or below, the deviation must be within ± 6%.
The three calibration points (20%, 60% & 100%) are selected by pressing .
NON ISO 6789-1:2017
Allows the user to set upper and lower limits from 0% to 99% of the target value.
TIP. For Indicating torque tools (Type I), the Dynatester is designed for class B, C and E tools.
Tools of class A and D should only be used with capacities of 10 N.m or below.
For Setting torque tools (Type II), the Dynatester is programmed for Class A, B and C
tools. Tools of class D, E, F and G should only be used with capacities of 10 N.m or
below.
The maximum target value in lbf.ft (ft lb) is 50 lbf.ft for the Dynatester 682/60 and 300 lbf.
ft for the Dynatester 682/400.
Operation
Measurement mode
Limit operation
Track
Limits follow signal and are not held.
Dial & Electronic
Limits status is held until ‘ZERO’ is pressed.
Click & Cam
Limit status is held until after the auto reset timer has operated.
Limits are enabled to work in both the Clockwise and Counter clockwise directions.
To change the limit target value, use or in any measurement mode.
Keep either key depressed for fast target value change.
Status indication
Torque signal
Display
LED’s
Serial port
Under lower limit
LO
Yellow
LO
Within limits
OK
Green
OK
Above upper limit
HI
Red
HI