8.2
Disassemble the Shield
Before decommissioning, it must be ensured that all sources are correctly inserted
into the shield and fixed, see chapter 7.5.1 Removing the Source, section
"Dismantle Shield".
1
2
3
4
Position
CLOSED
(lock slide or locking cylinder)
Padlock
Shield head
Handle bar (only slide lock variants)
Fig. 23
Dissasemble the shield
1.
Ensure that the shield has been closed and secured.
−
The locking cylinder is set to the
CLOSED
position (Fig. 23, item 1).
−
The padlock is installed (Fig. 23, item 2).
2.
Remove the handle bar (Fig. 23, item 4).
1 Lifting eye
Fig. 24
Attach shield and set down
3.
Attach the shield to the lifting eyes with adequately dimensioned and tested
slings (Fig. 24, item 1).
4.
Remove all screwed connections on the flange.
5.
Carefully remove the shield and carefully drop off the shield.
The decommissioning is completed.