Shields for Flange Mounting
3 System Description
44192BA2, Rev.03, 07/2021
17
The shield has the following functions:
•
Shield radiation to a level that is non-hazardous for the operating personnel
•
Locking the radiation beam outlet channel for transport and during installation
•
Protecting the integrated source capsule from mechanical damage and the
effects of the surrounding environment
3.1
Variant Locking Cylinder
1
2
3
4
5
Shield head with cover
Clamping screw
Coupling part of the source holder
Source holder
Source
6
7
8
9
Shield material
Locking cylinder (
CLOSED
position)
Housing
Locking lever
Fig. 3
Sectional view of a variant source with locking cylinder