6 - 2
BeneVision TM80 Telemetry Monitor Operator’s Manual
6.3
Understanding the Alarms
6.3.1 Alarm Categories
The telemetry monitor’s alarms can be classified into two categories: physiological alarms
and technical alarms
■
Physiological alarms, also called patient status alarms, are triggered by a monitored
parameter value that violates set alarm limits or an abnormal patient condition.
■
Technical alarms, also called system status alarms, are triggered by a device
malfunction or a patient data distortion due to improper operation or mechanical
problems.
Apart from the physiological and technical alarm messages, the telemetry monitor will
show some messages telling the system status in the prompt message area of the main
screen.
6.3.2 Alarm Priorities
Alarms can be classified into the following severity levels:
■
High priority alarms: indicates a life threatening situation or a severe device
malfunction. High priority alarms require an immediate response.
■
Medium priority alarms: indicates abnormal vital signs or a device malfunction.
Medium priority alarms require a timely response.
■
Low priority alarms: indicate a discomfort condition, a device malfunction, or an
improper operation. Low priority alarms require you to be aware of this condition.
■
Messages: provides additional information on the patient or the equipment.
6.3.3 Alarm Indicators
When an alarm occurs, the telemetry monitor indicates it through visual or audible alarm
indications. For more information, see the following table.
Alarm Indicator
High
Priority
Alarm
Medium
Priority
Alarm
Low
Priority
Alarm
Message
Comments
Alarm lamp
Red
Flashing
frequency:
1.4Hz -
2.8 Hz
Duty cycle:
20%-60%
Yellow
Flashing
frequency:
0.4Hz -0.8
Hz
Duty cycle:
20% -60%
Cyan
No
flashing
Duty
cycle:
100% on
None
None
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...