Commissioning and testing
40
TM800_D00165_00_M_XXEN/06.2015
5.3.1
Settings at the TM800
The following settings can optionally be carried out via the function buttons at the TM800 or the
TMK-SET software. Make a note of the modified settings (setting table).
Continue with chapter "Tests after parameter setting" on page 42.
Set language
Set BMS bus addresses
(int./ ext.)
Set alarm addresses and
system numbers
Set test addresses and
types of Isometers
TM800 with BMI8/8 resp.
BMI8/4 ?
Use of standard assignment
for medical gases/UPS
systems?
Can the buzzer sound be
differentiated from other
device signals?
Common
acknowledgement on/off
yes
Set time/date
Use of standard texts?
Use TMK-SET for
parameterisation (refer to
5.3.2)
no
Setting "medical"
Wiring acc. to specifications!
Relay: Set function and
switching behaviour
no
Adapt buzzer settings
Set and enable password
no
Relay or pushbutton
parameterisation?
yes
yes