Benchmark Scientific Incu-Shaker CO2 Mini Operating Manual Download Page 8

 

5. CO

  Incubator Operation 

 

5.1 Power switch 

Turn on the power switch. The digital LED will display current temperature and CO

% in the chamber. 

 

5.2 Setting temperature 
 

a. 

Press the "TEMP SET" key

, then, the LED screen will flash and display current programmed temperature.   

b. 

Set up the desired temperature by pressing 

UP (▲) 

or 

DOWN (▼).

 

c. 

Press "TEMP/SET" key

 after adjustment. The “SAVE” message is shown on the display.   

 
 

 

 

Note:  

Always press the “SET/TEMP” key after adjusting to the desired value.  Otherwise, the value will not be saved. 

 

5.3 Setting CO2 

 

a. 

Press "CO2 SET" key

. Then, the LED screen will begin flashing. 

b. 

Input the desired value of Co2 density

 by adjusting 

UP (▲) 

or 

DOWN (▼) key

 

c. 

Press "SET" key again

 after inputting the value. “SAVE” is shown up on LED screen as below.   

 

 

 

Note: 

Always press the “SET/CO2” key after adjusting to the desired value.  Otherwise, the value will not be saved.

 

   

 

5.4 Calibration for temperature and CO2

 

 

In the event of a discrepancy, please follow the below procedure for adjusting the calibration.   

 

Measure CO2 density and Temperature after incubator is stabilized.    (Allow 2 hours for stabilization after the 

desired setting has been reached.   

Please note that a very low deviation range such as 

±

0.1 ~ 0.3% could be difficult to correct precisely by this 

calibration.   

        

No. 

DISPLAY 

FUNCTION 

 

Chamber Temperature 

 

Door Heater Temperature 

 

Glass Door Heater Temperature 

 

CO2 Calibration 

 

 

Heating control 

 

Summary of Contents for Incu-Shaker CO2 Mini

Page 1: ...1 OPERATING MANUAL...

Page 2: ...Table of Contents NO CONTENTS PAGE 1 SAFETY AND GENERAL PRECAUTION 3 4 2 FEATURES AND SPECIFICATIONS 5 3 CONTROL PANEL 6 4 INSTALLATION 7 5 CO2 INCUBATOR OPERATION 8 11 6 ORBITAL SHAKER OPERATION 11 1...

Page 3: ...he event of smoke or a burning smell immediately remove the power cord from the outlet 1 3 Precaution for use Do not attempt to disassemble this instrument If service is required please contact your l...

Page 4: ...0cm from the back wall Install the unit in the way that the power plug is easily accessible and can be easily pulled in case of danger Install the instrument on a flat and stable surface free from vib...

Page 5: ...Microprocessor On Off Control Humidity 70 80 at 37 CO2 Range 0 to 20 Accuracy 0 1 at 5 at 37 setting CO2 Increment 0 1 CO2 Sensor IR CO2 Sensor Shaker RPM 30 to 300 rpm RPM Accuracy 1 Increment 1 RPM...

Page 6: ...6 Shaker Status Pilot Lamp When Shaker is running this lamp is flashing 7 Alarm On Off Switch 8 Alarm Mute Button When pressed provides a10 minutes audible alarm delay 9 Heating Signal Shows the stat...

Page 7: ...cold air Direct sunlight An uneven surface or table A place with heavy vibration A place with little air ventilation space behind the incubator 4 2 Place the shelves and water container as shown in ch...

Page 8: ...t the desired value of Co2 density by adjusting UP or DOWN key c Press SET key again after inputting the value SAVE is shown up on LED screen as below Note Always press the SET CO2 key after adjusting...

Page 9: ...th channel the LED is back to temperature display b Second Calibration Setting Outer Door Temperature The purpose of this heater is to prevent water condensation from forming on the glass door caused...

Page 10: ...ay shows 4 then press up to set to 1 e Fifth Calibration Setting Heating control NOTE Channel 5 is to set heating control point and is designed to prevent a significant temperature overshoot This mode...

Page 11: ...nces So set the temperature at 5 higher than operating temperature or desired point If the SAFETY LED on the front panel is illuminated this switch has been activated and is limiting the temperature 6...

Page 12: ...humb screws 6 4 Shaker Error Message 1 This message rErrL represents a low speed speed error Check the platform to ensure that nothing is obstructing it s movement Check the set speed to make sure tha...

Page 13: ...l number label of the machine 5 Defective power switch check voltage at switch 6 5 Disconnecting inside Shaker Connector of the orbital shaker is located at upper right back wall of chamber see below...