Benchmark Scientific Incu-Shaker CO2 Mini Operating Manual Download Page 6

 

3.Control Panel 

 
 
 

                 1               2                3      4            5                      6

 

 
 
 
 

 

 
 

        7        8     9     10     11

       12   13            14   15     16     17

 

 

1.  SAFETY activation indicator: If the over-temperature safety is activated, this LED will be ON 
2.  Temperature display LED window 

3.  CO

Gas % display window 

4.  CO

  Supply Status Pilot Lamp : Pilot lamp ON means inflow of gas 

5.  RPM display : RPM value display 
6.  Shaker Status Pilot Lamp : When Shaker is running, this lamp is flashing. 
7.  Alarm On/Off Switch   
8.  Alarm Mute Button : When pressed, provides a10 minutes audible alarm delay.     
9.  Heating Signal: Shows the status of each heating at the 3 sensor locations. (Glass door = heater around 

glass door.)   

10.  Temperature Setting Button. 
11.  Adjustment button 
12.  CO

Gas Setting Button 

13.  Calibration Mode button 
14.  Display Mode Selection Button either RPM or TIME 
15.  RPM setting Button 
16.  Adjustment button 
17.  Shaker Start/Stop Button   

 

 
 
 

      

 

 

 

Summary of Contents for Incu-Shaker CO2 Mini

Page 1: ...1 OPERATING MANUAL...

Page 2: ...Table of Contents NO CONTENTS PAGE 1 SAFETY AND GENERAL PRECAUTION 3 4 2 FEATURES AND SPECIFICATIONS 5 3 CONTROL PANEL 6 4 INSTALLATION 7 5 CO2 INCUBATOR OPERATION 8 11 6 ORBITAL SHAKER OPERATION 11 1...

Page 3: ...he event of smoke or a burning smell immediately remove the power cord from the outlet 1 3 Precaution for use Do not attempt to disassemble this instrument If service is required please contact your l...

Page 4: ...0cm from the back wall Install the unit in the way that the power plug is easily accessible and can be easily pulled in case of danger Install the instrument on a flat and stable surface free from vib...

Page 5: ...Microprocessor On Off Control Humidity 70 80 at 37 CO2 Range 0 to 20 Accuracy 0 1 at 5 at 37 setting CO2 Increment 0 1 CO2 Sensor IR CO2 Sensor Shaker RPM 30 to 300 rpm RPM Accuracy 1 Increment 1 RPM...

Page 6: ...6 Shaker Status Pilot Lamp When Shaker is running this lamp is flashing 7 Alarm On Off Switch 8 Alarm Mute Button When pressed provides a10 minutes audible alarm delay 9 Heating Signal Shows the stat...

Page 7: ...cold air Direct sunlight An uneven surface or table A place with heavy vibration A place with little air ventilation space behind the incubator 4 2 Place the shelves and water container as shown in ch...

Page 8: ...t the desired value of Co2 density by adjusting UP or DOWN key c Press SET key again after inputting the value SAVE is shown up on LED screen as below Note Always press the SET CO2 key after adjusting...

Page 9: ...th channel the LED is back to temperature display b Second Calibration Setting Outer Door Temperature The purpose of this heater is to prevent water condensation from forming on the glass door caused...

Page 10: ...ay shows 4 then press up to set to 1 e Fifth Calibration Setting Heating control NOTE Channel 5 is to set heating control point and is designed to prevent a significant temperature overshoot This mode...

Page 11: ...nces So set the temperature at 5 higher than operating temperature or desired point If the SAFETY LED on the front panel is illuminated this switch has been activated and is limiting the temperature 6...

Page 12: ...humb screws 6 4 Shaker Error Message 1 This message rErrL represents a low speed speed error Check the platform to ensure that nothing is obstructing it s movement Check the set speed to make sure tha...

Page 13: ...l number label of the machine 5 Defective power switch check voltage at switch 6 5 Disconnecting inside Shaker Connector of the orbital shaker is located at upper right back wall of chamber see below...

Reviews: